Endoscopy 2005; 37(6): 602
DOI: 10.1055/s-2005-861100
Unusual Cases and Technical Notes
© Georg Thieme Verlag KG Stuttgart · New York

A Percutaneous Endoscopic Gastrostomy Tube Pushed Through the Posterior Wall of the Stomach as far as the Pancreas During Nonendoscopic Tube Exchange

S. Kasamaki1, 2 , T. Kamano2 , Y. Hayashida3 , H. Motoyama1 , H. Yokota1
  • 1Department of Surgery, Hokota Hospital, Ibaraki Prefecture, Japan
  • 2Department of Coloproctological Surgery, Juntendo University, School of Medicine, Tokyo, Japan
  • 3Department of General Medicine, Juntendo University, School of Medicine, Tokyo, Japan
Further Information

Publication History

Publication Date:
16 May 2006 (online)

The safety and efficacy of percutaneous endoscopic gastrostomy (PEG) have been well documented [1] [2]. With regard to the removal of PEG tubes, the standard method used recently has been nonendoscopic removal: the percutaneous removal method should cost less and be less time-consuming because it obviates the need for an endoscopy. However, complications arising from such PEG tube exchange procedures have been reported [3] [4]. We encountered a patient who developed problems as a result of a PEG tube exchange, in whom it was discovered that the the PEG tube had gone through the posterior wall of the stomach, reaching as far as the anterior aspect of the pancreas.

A 78-year-old woman underwent the insertion of a 16-Fr Foley catheter using a stylet under pressure instead of a new PEG tube because the insertion of a new tube was difficult. Sudden-onset abdominal pain and a fever of over 37.9 ° C occurred as soon as enteral feeding was commenced and she was transferred to our hospital. Endoscopic examination revealed that the Foley catheter had gone through the posterior wall of the stomach (Figure [1]). Computed tomography revealed that the tip of the catheter had reached the anterior aspect of the pancreas (Figure [2]). After deflating the balloon by 5 ml per week, a second endoscopic examination revealed converging folds in the posterior wall of the stomach. A new PEG tube was inserted endoscopically using the pull technique.

Figure 1 Endoscopic examination revealed that the Foley catheter had gone through the posterior wall of the stomach.

Figure 2 Computed tomography revealed that the Foley catheter’s balloon had gone through the posterior wall of the stomach and reached the front of the pancreas.

Mechanical stimulation of the mucosa of the stomach by the tip of the PEG tube has been reported to lead to the development of gastric ulcer [5]. However, the gastric ulcer in our case was caused by the excessive force used to insert the new PEG tube and this was unacceptable. Endoscopists skilled in PEG techniques should carry out PEG tube exchanges in order to avoid malpositioning of PEG tubes.

Endoscopy_UCTN_Code_CPL_1AH_2AI

References

  • 1 Ponsky J L. Percutaneous endoscopic gastrostomy: techniques of removal and replacement.  Gastrointest Endosc Clin N Am. 1992;  2 215-221
  • 2 Stephen A M, Wei-Kuo C. Complications of enteral access.  Gastrointest Endosc. 2003;  58 739-751
  • 3 Cass O W, Rowland K, Bartram B. et al . Insertion, efficacy, and removal of a nonendoscopically removable percutaneous endoscopic gastrostomy (PEG) tube.  Surg Endosc. 1999;  13 516-519
  • 4 Vu C K. Buried bumper syndrome: old problem, new tricks.  J Gastroenterol Hepatol. 2002;  17 1125-1128
  • 5 Kanie J, Akatsu H, Suzuki Y. et al . Mechanism of the development of gastric ulcer after percutaneous endoscopic gastrostomy.  Endoscopy. 2002;  34 480-482

S. Kasamaki, M. D.

Department of Coloproctological Surgery, Juntendo University, School of Medicine

Hongo 2-1-1
Bunkyo-Ku
Tokyo 113-8421
Japan

Fax: +81-3-3812-1056

Email: kasamaki@hh.iij4u.or.jp

    >