Abstract
Shoulder massage machines have two pads that are driven by solenoid coils to perform a per cussive massage on the shoulders. There have been concerns that such machines might create electromagnetic interference (EMI) in implantable cardiac devices because of the time-varying magnetic fields produced by the alternating current in the solenoid coils. The objective of this study was to investigate the potential EMI from one such shoulder massage machine on implantable cardiac devices. We measured the distribution profile of the magnetic field intensity around the massage machine. Furthermore, we performed an inhibition test and an asynchronous test on an implantable cardiac pacemaker using the standardized Irnich human body model. We examined the events on an implantable cardioverter–defibrillator (ICD) using a pacemaker programmer while the massage machine was in operation. The magnetic field distribution profile exhibited a peak intensity of 212 (A/m) in one of the solenoid coils. The maximal interference distance between the massage machine and the implantable cardiac pacemaker was 28 cm. Ventricular fibrillation was induced when the massage machine was brought near the electrode of the ICD and touched the Irnich human body model. It is necessary to provide a “don’t use” warning on the box or the exterior of the massage machines or in the user manuals and to caution patients with implanted pacemakers about the dangers and appropriate usage of massage machines.
Similar content being viewed by others
References
Hirose M, Hida M, Sato E, et al. Electromagnetic interference of implantable unipolar cardiac pacemakers by an induction oven. PACE. 2005;28:540–8.
Yamamoto T, Matsumoto K, Saitou J, et al. The effect of electromagnetic interference in pacemaker implanted patients by cellular phones. J Arrhythm. 1998;14:228–94.
Ministry of Internal Affairs and Communications (MIC) in Japan: Surveillance study report of Influence of the medical equipment on an electric wave. 2002.
Irnich W, Batz L, Muller R, et al. Electromagnetic interference of pacemaker by mobile phones. PACE. 1996;19:1431–46.
Gimble JR, Cox JW. Electronic article surveillance systems and interactions with implantable cardiac devices: risk of adverse interactions in public and commercial spaces. Mayo Clin Proc. 2007;82:318–22.
Toyoshima T. Electromagnetic interference for implantable device patients. Medical Review Co. Ltd. 2007. pp. 13–15 (Authors’ translation from Japanese).
Electromagnetic Compatibility Conference (EMCC). A glossary of EMC, Ohmsha. Ltd. 1999. pp. 122–123 (Authors’ translation from Japanese).
Toyoshima T. Electromagnetic interference for implantable device patients. Medical Review Co. Ltd. 2007. pp. 56 (Authors’ translation from Japanese).
Irnich W. Interference in pacemaker. PACE. 1984;7:1021–48.
ISO 14708(2005) Implants for surgery—Active implantable medical devices—Annex F.
Ministry of Internal Affairs and Communications (MIC) in Japan: Surveillance study report of Influence of the medical equipment on an electric wave. 2011.
Van Gelder BM, Van Den Broek W, et al. Paradoxical atrial undersensing: noise rate reversion or amplifier ringing? J Cardiovasc Electrophysiol. 2006;17:1371–4.
Toyoshima T, Tsumura M, Nojima T, et al. Electromagnetic interference of implantable cardiac pacemaker by portable telephones. Card Pacing. 1996;12:488–97.
Acknowledgments
The authors would like to thank Hiroshi Fujimoto and Katsumi Yokomizo of Medtronic Japan Co. Ltd. for loaning us the implantable cardiac devices used in this study and for offering us their advice.
Conflict of interest
The authors declare that they have no conflict of interest.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Yoshida, S., Fujiwara, K., Kohira, S. et al. Electromagnetic interference of implantable cardiac devices from a shoulder massage machine. J Artif Organs 17, 243–249 (2014). https://doi.org/10.1007/s10047-014-0765-1
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10047-014-0765-1