Abstract
Background
Reliable laparoscopic fixation of meshes prior to their fibrous incorporation is intended to minimize recurrences following transabdominal preperitoneal hernia repair (TAPP) and totally extraperitoneal repair (TEP) repair of inguinal hernias. However, suture-, tack- and staple-based fixation systems are associated with postoperative chronic inguinal pain. Initial fixation with fibrin sealant offers an atraumatic alternative, but there is little data demonstrating directly whether fibrin-based mesh adhesion provides adequate biomechanical stability for repair of inguinal hernia by TAPP and TEP.
Methods
Using a newly developed, standardized simulation model for abdominal wall hernias, sublay repairs were performed with six different types of commercially available hernia mesh. The biomechanical stability achieved, and the protection afforded by the mesh–hernia overlap, were compared for three different techniques: nonfixation, point-by-point suture fixation, and fibrin sealant fixation.
Results
Mesh dislocation from the repaired hernia defect was consistently seen with nonfixation. This was reliably prevented with all six mesh types when fixed using either sutures or fibrin sealant. The highest stress resistance across the whole abdominal wall was found following superficial fixation with fibrin sealant across the mesh types. There was a highly statistically significant improvement in fixation stability with fibrin sealant versus fixation using eight single sutures (p = 0.008), as assessed by the range of achievable peak pressure stress up to 200 mmHg.
Conclusions
To ensure long-term freedom from recurrence, intraoperative mesh–hernia overlap must be retained. This can be achieved with fibrin sealant up to the incorporation of the mesh – without trauma and with biomechanical stability.
Similar content being viewed by others
References
Mohr D, Bauer J, Döbler K, Fischer B, Woldenga C (2003) BQS-Qualitätsbericht 2002- Modul 12/3: Hernienoperation. Bundesgeschäftsstelle Qualitätssicherung gGmbH, Düsseldorf
McCormack K, Scott NW, Go PM, Ross S, Grant AM - EU Hernia Trialists Collaboration (2003) Laparoscopic techniques versus open techniques for inguinal hernia repair. Cochrane Database Syst Rev 1: CD001785
EU Hernia Trialists Collaboration (2002) Laparoscopic versus open groin hernia repair: meta-analysis of randomised trials based on individual patient data. Hernia 6:2–10
Beattie GC, Rumar S, Nixon SJ (2000) Laparoscopic total extraperitoneal hernia repair: mesh fixation is unnecessary. J Laparoendosc Adv Surg 10:71–73
Poobalan AS, Bruce J, Cairns W, Smith S, King PM, Krukowski ZH, Chambers W (2003) A review of chronic pain after inguinal herniorrhaphy. Clin J Pain 19:48–54
Stark E, Oestreich K, Wendl K, Rumstadt B, Hagmüller E (1999) Nerve irritation after laparoscopic hernia repair. Surg Endosc 13:878–881
Neumayer L, Giobbie-Hurder A, Jonasson O, Fitzgibbons R Jr, Dunlop D, Gibbs J, Reda D, Henderson W, Veterans Affairs Cooperative Studies Program 456 Investigators (2004) Open mesh versus laparoscopic mesh repair of inguinal hernia. N Engl J Med 350:1819–1827
Kumar S, Wilson RG, Nixon SJ, Macintyre IMC (2002) Chronic pain after laparoscopic and open mesh repair of groin hernia. Br J Surg 89:1476–1479
Hindmarsh AC, Cheong E, Lewis MPN, Rhodes M (2003) Attendance at a pain clinic with severe chronic pain after open and laparoscopic inguinal hernia repairs. Br J Surg 90:1152–1154
Callesen T, Bech K, Kehlet H (1999) Prospective study of chronic pain after hernia repair. Br J Surg 86:1528–1531
Ferzli GS, Frezza E, Pecoraro AM, Dee Ahern K (1999) Prospective randomized study of stapled versus unstapled mesh in a laparoscopic preperitoneal inguinal hernia repair. J Am Coll Surg 1881:461–465
Lau H, Nivritti GP (2003) Selective non-stapling of mesh during unilateral endoscopic extraperitoneal inguinal hernioplasty. Arch Surg 138:1352–1355
Smith AI, Royston CM, Sedman PC (1999) Stapled and non-stapled laparoscopic transabdominal TAPP inguinal hernia repair: a prospective randomized trial. Surg Endosc 13:804–806
Khajanchee Y, Urbach R, Swanstrom L, Hansen P (2001) Outcomes of laparoscopic herniorrhaphy without fixation of mesh to the abdominal wall. Surg Endosc 15:1102–1107
Chevrel JP, Rath AM (1997) The use of fibrin glues in the surgical treatment of incisional hernias. Hernia 1:9–14
Katkhouda N, Mavor E, Friedlander MH, Mason RJ, Kiyabu M, Grant SW, Achanta K, Kirkman EL, Narayanan K, Essani R (2001) Use of fibrin sealant for prosthetic mesh fixation in laparoscopic extraperitoneal inguinal hernia repair. Ann Surg 233:18–25
Topart P, Vandenbroucke F, Lozac’h P (2005) Tisseel vs tack staples as mesh fixation in totally extraperitoneal laparoscopic repair of groin hernias. Surg Endosc 19:724–727
Lau H (2005) Fibrin sealant versus mechanical stapling for mesh fixation during endoscopic extraperitoneal inguinal hernioplasty: a randomized prospective trial. Ann Surg 242:670–675
Schwab R, Willms A, Kröger A, Becker HP (2006) Less chronic pain following mesh fixation using fibrin sealant in TEP inguinal hernia repair. Hernia 10:272–277
Novik B, Hagdorn S, Mörk UB, Dahlin K, Skullmann S, Dahlenbäck J (2006) Fibrin glue for securing the mesh in laparoscopic totally extraperitoneal inguinal hernia repair: a study with a 40-month prospective follow-up period. Surg Endosc 20:462–467
Junge K, Klinge U, Prescher A, Giboni P, Niewiera M, Schumpelick V (2001) Elasticity of the anterior abdominal wall and impact for reparation of incisional hernias using mesh implants. Hernia 5:113–118
Lowham AS, Filipi CJ, Fitzgibbons RJ, Stoppa R, Wantz GE, Felix EL, Crafton WB (1997) Mechanisms of hernia recurrence after preperitoneal mesh repair. Traditional and laparoscopic. Ann Surg 225:422–431
Dion YM, Charara J, Guidoin R (1994) Bursting strength evaluation. Comparison of 0 - Prolene sutures and endoscopic staples in an experimental prosthetic patch repair of abdominal wall defect. Surg Endosc 8:812–816
Brockman JB, Patterson NW, Richardson WS (2004) Burst strength of laparoscopic and open hernia repair. Surg Endosc 18:536–539
Hollinsky C, Goebl S (1999) Bursting strength evaluation after different types of mesh fixation in laparoscopic herniorrhaphy. Surg Endosc 13:958–961
Horgan LF, Shelton JC, O’Riordan DC, Moore DP, Winslet MC, Davidson BR (1996) Strengths and weaknesses of laparoscopic and open mesh inguinal hernia repair: a randomized controlled experimental study. Br J Surg 83:1463–1467
Kes E, Lange J, Bonjer J, Stoeckart R, Mulder P, Snijders C, Kleinrensink G (2004) Protrusion of prosthetic meshes in repair of inguinal hernias. Surgery 135(2):163–70
Knook MT, van Rosmalen AC, Yoder BE, Kleinrensink GJ, Snijders CJ, Looman CW, van Steensel CJ (2001) Optimal mesh size for endoscopic inguinal hernia repair: a study in a porcine model. Surg Endosc 15:1471–1477
Song C, Alijani A, Frank T, Hanna GB, Cuschieri A (2006) Mechanical properties of the human abdominal wall measured in vivo during insufflation for laparoscopic surgery. Surg Endosc 20:987–990
Zieren J, Castenholz E, Jacobi CA, Zieren HU, Müller J (1999) Is mesh fixation in abdominal hernia necessary? Langenbecks Arch Surg 384:71–75
Petter-Puchner AH, Fortelny R, Mittermayr R, Öhlinger W, Redl H (2005) Fibrin sealing versus stapling of hernia meshes in an onlay model in the rat. Hernia 9:322–329
van’t Riet M, de Vos van Steenwijk PJ, Kleinrensink GJ, Steyerberg EW, Bonjer HJ (2002) Tensile strength of mesh fixation methods in laparoscopic incisional hernia repair. Surg Endosc 16:1713–1716
Klinge U, Schumpelick V (2003) Prosthetic implants for hernia repair. Br J Surg 90:1457–1458
Junge K, Peiper C, Schachtrupp A, Rosch R, Kürten D, Klinge U, Schumpelick V (2003) Breaking strength and tissue elasticity after Shouldice repair. Hernia 7:17–20
Acknowledgements
This work was supported by the Baxter AG, Heidelberg, Germany, who provided the tissue glue. Furthermore, we are grateful to Mr. Ulf Boerner and the Fraunhofer Institute for Production Technologies (IPT), (Aachen, Germany) for excellent and careful assistance during the development of the hernia test stand.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Schwab, R., Schumacher, O., Junge, K. et al. Biomechanical analyses of mesh fixation in TAPP and TEP hernia repair. Surg Endosc 22, 731–738 (2008). https://doi.org/10.1007/s00464-007-9476-5
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00464-007-9476-5