Real-life performance of swept-source optical coherence tomography angiography

Experiences with an evolving technique

verfasst von: Gerasimos Anastassiou, Abdulmonem Elmajie, Claudia Smok, Stephan Kremmer

Erschienen in: Spektrum der Augenheilkunde | Ausgabe 6/2017

Einloggen, um Zugang zu erhalten

Summary

Background

Optical coherence tomography (OCT) angiography (OCTA) is a novel technology for imaging the retinal vasculature. The real-life applicability of OCTA was assessed in a busy outpatient eye clinic using the same hardware (Triton, TOPCON, Tokyo, Japan) but different software versions 1 year apart. The influence of various parameters on OCTA performance was assessed.

Materials and methods

Two groups of 100 patients with intravitreal injections for maculopathy were evaluated with OCTA. Demographic characteristics did not differ significantly. Group A was assessed in February 2016, group B in February 2017. Recording time, scan quality, visual acuity, patient age, diagnosis, and lens status were recorded for statistical evaluation.

Results

The time taken for examination differed significantly between the two groups. In group A, average recording time was 44.7 s (standard deviation, SD, 26.9 s; range 10–103 s), in group B 17.1 s (SD 9.0 s, range 7–53 s; P = 0.001). Image quality was higher in group B (P = 0.03). In bivariate analysis, recording time, age, and visual acuity correlated with image quality. In univariate regression analysis, recording duration and visual acuity were associated with a higher risk of poor OCTA quality. In multivariate regression analysis, scan duration was the sole predictor of a higher risk for poor OCTA quality (group A: odds ratio, OR, 5.9; confidence interval, CI, 2.0–17.5; P = 0.001; group B: OR 3.9, CI 1.3–11.9, P = 0.01).

Conclusion

Advances in software-based artefact correction reduce recording time and improve SS-OCTA quality significantly. Although SS-OCTA is still more time consuming than conventional SS-OCT, current performance enables routine application even in a busy retina outpatient clinic. Further hardware and software improvements should focus on recording time.
Literatur
1.
Zurück zum Zitat Told R, Ginner L, Hecht A, Sacu S, Leitgeb R, Pollreisz A, et al. Comparative study between a spectral domain and a high-speed single-beam swept source OCTA system for identifying choroidal neovascularization in AMD. Sci Rep. 2016;6:38132. doi:10.1038/srep38132.CrossRefPubMedPubMedCentral Told R, Ginner L, Hecht A, Sacu S, Leitgeb R, Pollreisz A, et al. Comparative study between a spectral domain and a high-speed single-beam swept source OCTA system for identifying choroidal neovascularization in AMD. Sci Rep. 2016;6:38132. doi:10.​1038/​srep38132.CrossRefPubMedPubMedCentral
2.
Zurück zum Zitat Zhang Q, Chen CL, Chu Z, Zheng F, Miller A, Roisman L, et al. Automated quantitation of choroidal neovascularization: a comparison study between spectral-domain and swept-source OCT angiograms. Invest Ophthalmol Vis Sci. 2017;58(3):1506–13.CrossRefPubMedPubMedCentral Zhang Q, Chen CL, Chu Z, Zheng F, Miller A, Roisman L, et al. Automated quantitation of choroidal neovascularization: a comparison study between spectral-domain and swept-source OCT angiograms. Invest Ophthalmol Vis Sci. 2017;58(3):1506–13.CrossRefPubMedPubMedCentral
3.
Zurück zum Zitat de Carlo TE, Chin AT, Bonini Filho MA, Adhi M, Branchini L, Salz DA, et al. Detection of microvascular changes in eyes of patients with diabetes but not clinical diabetic retinopathy using optical coherence tomography angiography. Retina (Philadelphia, Pa). 2015;35(11):2364–70.CrossRef de Carlo TE, Chin AT, Bonini Filho MA, Adhi M, Branchini L, Salz DA, et al. Detection of microvascular changes in eyes of patients with diabetes but not clinical diabetic retinopathy using optical coherence tomography angiography. Retina (Philadelphia, Pa). 2015;35(11):2364–70.CrossRef
5.
Zurück zum Zitat Kashani AH, Lee SY, Moshfeghi A, Durbin MK, Puliafito CA. Optical coherence tomography angiography of retinal venous occlusion. Retina (Philadelphia, Pa). 2015;35(11):2323–31.CrossRef Kashani AH, Lee SY, Moshfeghi A, Durbin MK, Puliafito CA. Optical coherence tomography angiography of retinal venous occlusion. Retina (Philadelphia, Pa). 2015;35(11):2323–31.CrossRef
6.
Zurück zum Zitat Spaide RF, Klancnik JM Jr, Cooney MJ. Retinal vascular layers imaged by fluorescein angiography and optical coherence tomography angiography. JAMA Ophthalmol. 2015;133(1):45–50.CrossRefPubMed Spaide RF, Klancnik JM Jr, Cooney MJ. Retinal vascular layers imaged by fluorescein angiography and optical coherence tomography angiography. JAMA Ophthalmol. 2015;133(1):45–50.CrossRefPubMed
7.
Zurück zum Zitat Kremmer S, Smok C, Anastassiou G. Swept-Source OCT (SS-OCT) and Swept Source OCT Angiography (SS-OCTA) in the Diagnostics of Glaucoma. Spektrum Augenheilkd 2017;31. doi:10.1007/s00717-017-0372-7 Kremmer S, Smok C, Anastassiou G. Swept-Source OCT (SS-OCT) and Swept Source OCT Angiography (SS-OCTA) in the Diagnostics of Glaucoma. Spektrum Augenheilkd 2017;31. doi:10.​1007/​s00717-017-0372-7
8.
Zurück zum Zitat Akil H, Al-Sheikh M, Falavarjani KG, Francis B, Chopra V. Choroidal thickness and structural glaucoma parameters in glaucomatous, preperimetric glaucomatous, and healthy eyes using swept-source OCT. Eur J Ophthalmol. 2017; doi:10.5301/ejo.5000926.PubMed Akil H, Al-Sheikh M, Falavarjani KG, Francis B, Chopra V. Choroidal thickness and structural glaucoma parameters in glaucomatous, preperimetric glaucomatous, and healthy eyes using swept-source OCT. Eur J Ophthalmol. 2017; doi:10.​5301/​ejo.​5000926.PubMed
9.
Zurück zum Zitat Salz DA, de Carlo TE, Adhi M, Moult E, Choi W, Baumal CR, et al. Select features of diabetic retinopathy on swept-source optical coherence tomographic angiography compared with fluorescein angiography and normal eyes. JAMA Ophthalmol. 2016;134(6):644–50.CrossRefPubMedPubMedCentral Salz DA, de Carlo TE, Adhi M, Moult E, Choi W, Baumal CR, et al. Select features of diabetic retinopathy on swept-source optical coherence tomographic angiography compared with fluorescein angiography and normal eyes. JAMA Ophthalmol. 2016;134(6):644–50.CrossRefPubMedPubMedCentral
10.
Zurück zum Zitat Spaide RF, Fujimoto JG, Waheed NK. Image artifacts in optical coherence tomography angiography. Retina (Philadelphia, Pa). 2015;35(11):2163–80.CrossRef Spaide RF, Fujimoto JG, Waheed NK. Image artifacts in optical coherence tomography angiography. Retina (Philadelphia, Pa). 2015;35(11):2163–80.CrossRef
11.
Zurück zum Zitat Ghasemi Falavarjani K, Al-Sheikh M, Akil H, Sadda SR. Image artefacts in swept-source optical coherence tomography angiography. Br J Ophthalmol. 2017;101(5):564–8.CrossRefPubMed Ghasemi Falavarjani K, Al-Sheikh M, Akil H, Sadda SR. Image artefacts in swept-source optical coherence tomography angiography. Br J Ophthalmol. 2017;101(5):564–8.CrossRefPubMed
Metadaten
Titel
Real-life performance of swept-source optical coherence tomography angiography
Experiences with an evolving technique
verfasst von
Gerasimos Anastassiou
Abdulmonem Elmajie
Claudia Smok
Stephan Kremmer
Publikationsdatum
23.08.2017
Verlag
Springer Vienna
Erschienen in
Spektrum der Augenheilkunde / Ausgabe 6/2017
Print ISSN: 0930-4282
Elektronische ISSN: 1613-7523
DOI
https://doi.org/10.1007/s00717-017-0361-x

Weitere Artikel der Ausgabe 6/2017

Spektrum der Augenheilkunde 6/2017 Zur Ausgabe