Thorac Cardiovasc Surg 1989; 37(1): 47-51
DOI: 10.1055/s-2007-1013904
© Georg Thieme Verlag Stuttgart · New York

Lethal Blockage of a Bjørk-Shiley Artificial Heart Valve Caused by Strut Fracture - the Metallurgical Aspect

Blockierung einer Bjørk-Shiley-Klappenscheibe infolge Bügelbruchs mit tödlichem Ausgang - eine metallurgische UntersuchungG. Röckelein1 , J. Breme2 , J. von der Emde3
  • 1Pathological Institute
  • 2Materials Technology (Metals) Chair
  • 3Heart Surgery Department of the Surgical Clinic Erlangen-Nürnberg University, FRG
Further Information

Publication History

1988

Publication Date:
19 March 2008 (online)

Summary

The fracture of the outlet strut of a Bjørk-Shiley artificial heart valve implanted in a 21 year old man lead - 6 years after the operation - to the valve disc jamming closed with a resulting lethal heart failure and circulation standstill. The probable cause of the valve strut fracture could be determined by metallurgical/metallographical investigation of the material and the production methods.

Zusammenfassung

Der Bruch eines Metallbügels einer Björk-Shiley Prothese bei einem 21jährigen Mann führte 6 Jahre nach der Operation zum Arretieren der Prothesenscheibe in Schlußposition mit der Folge eines letalen Herzkreislaufstillstandes. Durch die werkstoffwissenschaftliche Untersuchung konnte die wahrscheinliche Ursache des Klappenstützenbruches in Grundwerkstoff und Verarbeitungsmethoden nachgewiesen werden.

    >