Contamination with Hepatitis B Virus DNA in Gastrointestinal Endoscope Channels: Risk of Infection on Reuse after On-Site Cleaning

Y.  Ishino1 , K.  Ido1 , K.  Sugano1
  • 1Dept. of Gastroenterology, Jichi Medical School, Tochigi, Japan
Further Information

Publication History

Submitted 21 March 2004

Accepted after Revision 10 December 2004

Publication Date:
03 June 2005 (online)

Abstract

Background and Study Aims: The incidence of viral contamination in the air, water and suction/accessory channels of gastrointestinal endoscopes was examined in order to evaluate the risk of infection.
Materials and Methods: After endoscopic examinations, including biopsy procedures, in 17 patients who were positive for hepatitis B virus surface antigen and eight patients who were positive for hepatitis C virus antibody, the endoscopes were cleaned on site by suctioning and flushing the air and water channels with an enzyme detergent. First samples were then collected by flushing 5 ml of sterile water through each channel. After mechanical reprocessing, second samples were collected in the same way. Virological studies were carried out with real-time polymerase chain reactions for hepatitis B virus DNA and hepatitis C virus RNA.
Results: Hepatitis B virus DNA was detected in five of the first samples recovered from the suction/accessory channels of the endoscopes (titers of 1.3 × 104 to 2.5 × 105 copies/ml), while no contamination was detected after reprocessing (P = 0.0445). The first samples from one water channel and three air channels were also positive for hepatitis B virus DNA, but were negative after reprocessing (P > 0.5, P = 0.227, respectively). No hepatitis C virus RNA was detected in any of the samples.
Conclusions: These results indicate that all of the channels were potential sources of viral infection.

References

  • 1 Lynch D A, Porter C, Murphy L. et al . Evaluation of four commercial automatic endoscope washing machines.  Endoscopy. 1992;  24 766-770
  • 2 Ido K, Ishino Y, Ota Y. et al . Deficiencies of automatic endoscopic reprocessors: a method to achieve high-grade disinfection of endoscopes.  Gastrointest Endosc. 1996;  44 583-586
  • 3 Urayama S, Kozarek R A, Sumida S. et al . Mycobacteria and glutaraldehyde: is high-level disinfection of endoscopes possible?.  Gastrointest Endosc. 1996;  43 451-456
  • 4 Tremain S C, Orientale E, Rodney W M. Cleaning, disinfection, and sterilization of gastrointestinal endoscopes: approaches in the office.  J Fam Pract. 1991;  32 300-305
  • 5 Fraser V J, Zuckerman G, Clouse R E. et al . A prospective randomized trial comparing manual and automated endoscope disinfection methods.  Infect Control Hosp Epidemiol. 1993;  14 383-389
  • 6 Ishino Y, Ido K, Koiwai H. et al . Pitfalls in endoscope reprocessing: brushing of air and water channels is mandatory for high-level disinfection.  Gastrointest Endosc. 2001;  53 165-168
  • 7 Alvarado C J, Reichelderfer M. APIC guideline for infection prevention and control in flexible endoscopy. Association for Professionals in Infection Control.  Am J Infect Control. 2000;  28 138-155
  • 8 Ogoshi K, Akamatsu T, Iishi H. et al . JGES guideline for cleaning and disinfection in gastrointestinal endoscopy.  Dig Endosc. 2000;  12 369-382
  • 9 Systchenko R, Marchetti B, Canard J N. et al . Guidelines of the French Society of Digestive Endoscopy: recommendations for setting up cleaning and disinfection procedures in gastrointestinal endoscopy.  Endoscopy. 2000;  32 807-818
  • 10 BSG E ndoscopy. Cleaning and disinfection of equipment for gastrointestinal endoscopy: report of a working party of the British Society of Gastroenterology endoscopy committee.  Gut. 1998;  42 585-593
  • 11 Rey J, Kruse A, Axon T. et al . ESGE guidelines for the prevention of endoscopic transmission of type C hepatitis and update on Creutzfeldt-Jakob disease.  Endoscopy. 1997;  29 203-204
  • 12 Kovacs B J, Chen Y K, Kettering J D. et al . High-level disinfection of gastrointestinal endoscopes: are current guidelines adequate?.  Am J Gastroenterol. 1999;  94 1546-1550
  • 13 Foliente R L, Kovacs B J, Aprecio R M. et al . Efficacy of high-level disinfectants for reprocessing GI endoscopes in simulated-use testing.  Gastrointest Endosc. 2001;  53 456-462
  • 14 Birnie G G, Quigley E M, Clements G B. et al . Endoscopic transmission of hepatitis B virus.  Gut. 1983;  24 171-174
  • 15 Kasugai T, Yoshii Y, Shikata J. et al . Digestive endoscopy and HBV infection (first report) [in Japanese with English abstract].  Gastroenterol Endosc. 1984;  27 2727-2733
  • 16 Bronowicki J P, Venard V, Botté C. et al . Patient-to-patient transmission of hepatitis C virus during colonoscopy.  N Engl J Med. 1997;  337 237-240
  • 17 Tytgat G N. Endoscopic transmission of Helicobacter pylori. Aliment Pharmacol Ther 1995 9 (Suppl 2): 105-110
  • 18 Akamatsu T, Tabata K, Hironga M. et al . Transmission of Helicobacter pylori infection via flexible fiberoptic endoscopy.  Am J Infect Control. 1996;  24 396-401
  • 19 Bond W W, Moncada R E. Viral hepatitis B infection risk in flexible fiberoptic endoscopy.  Gastrointest Endosc. 1978;  24 225-230
  • 20 Chanzy B, Duc-Bin D L, Rousset B. et al . Effectiveness of a manual disinfection procedure in eliminating hepatitis C virus from experimentally contaminated endoscopes.  Gastrointest Endosc. 1999;  50 147-151
  • 21 Abe A, Inoue K, Tanaka T. et al . Quantitation of hepatitis B virus genomic DNA by real-time detection PCR.  J Clin Microbiol. 1999;  37 2899-2903
  • 22 Takeuchi T, Katsume A, Tanaka T. et al . Real-time detection system for quantification of hepatitis C virus genome.  Gastroenterology. 1999;  116 636-642
  • 23 Deva A K, Vickery K, Zou J. et al . Establishment of an in-use testing method for evaluating disinfection of surgical instruments using the duck hepatitis B model.  J Hosp Infect. 1996;  33 119-130
  • 24 Vesley D, Melson J, Stanley P. Microbial bioburden in endoscope reprocessing and an in-use evaluation of the high-level disinfection capabilities of Cidex PA.  Gastroenterol Nurs. 1999;  22 63-68
  • 25 Deva A K, Vickery K, Zou J. et al . Detection of persistent vegetative bacteria and amplified viral nucleic acid from in-use testing of gastrointestinal endoscopes.  J Hosp Infect. 1998;  39 149-157
  • 26 Alfa M J, Jackson M. A new hydrogen peroxide-based medical-device detergent with germicidal properties: comparison with enzymatic cleaners.  Am J Infect Control. 2001;  29 168-177
  • 27 Di Marino A JJ. Noncompliance with FDA and society guidelines for endoscopic reprocessing: implications for patient care.  Gastrointest Endosc Clin N Am. 2000;  10 283-294
  • 28 Orsi G B, Filocamo A, di Stefano L. et al . Italian national survey of digestive endoscopy disinfection procedures.  Endoscopy. 1997;  29 732-740
  • 29 Cheung R J, Ortiz D, Di Marino A JJ. GI endoscopic reprocessing practices in the United States.  Gastrointest Endosc. 1999;  50 362-368
  • 30 Cleaning and Disinfection Committee for Endoscopes, Japanese Gastroenterological Endoscopy Society . The Cleaning and Disinfection Committee for Endoscopes of the Japanese Gastroenterological Endoscopy Society report: questionnaire survey of digestive endoscopy disinfection procedure [in Japanese].  Gastroenterol Endosc. 1999;  41 215-219

Y. Ishino, M.D.

Dept. of Gastroenterology · Jichi Medical School

Minamikawachi-machi · Kawachi Gun · Tochigi 329-0498 · Japan

Fax: + 81-285-44-8297

Email: ishino@jichi.ac.jp

    >