Endosc Int Open 2016; 04(01): E79-E82
DOI: 10.1055/s-0034-1393583
Original article
© Georg Thieme Verlag KG Stuttgart · New York

Anatomy-shaped design of a fully-covered, biliary, self-expandable metal stent for treatment of benign distal biliary strictures

Jochen Weigt
Department of Gastroenterology, Hepatology and Infectious Diseases, Otto-von-Guericke University, Magdeburg, Germany
,
Arne Kandulski
Department of Gastroenterology, Hepatology and Infectious Diseases, Otto-von-Guericke University, Magdeburg, Germany
,
Peter Malfertheiner
Department of Gastroenterology, Hepatology and Infectious Diseases, Otto-von-Guericke University, Magdeburg, Germany
› Author Affiliations
Further Information

Publication History

Publication Date:
27 November 2015 (online)

Background and study aims: The treatment success of benign biliary strictures with fully covered metal stents (CSEMS) is altered by high stent dislocation rates. We aimed to evaluate a new stent design to prevent dislocation.

Patients and methods: Patients with benign biliary strictures were treated with a newly designed double-coned stent (dcSEMS). Mechanical analysis of the new stent was performed and it was compared with a cylindrical stent.

Results: A total of 13 dcCSEMS were implanted in 11 patients (2 female, 9 male, median age 47, range 33 – 71). All patients had distal biliary strictures due to chronic pancreatitis. No stent migration occurred. In all but one patient the stents were removed. One patient refused stent extraction and was lost to follow up. Stent occlusion occurred twice leading to cholangitis in both cases. The duration of stent treatment was 170 days (range 61 – 254). After extraction only one patient had early recurrent stricture and received the same stent again. Three stents showed minimal tissue granulation at the papilla. One stent presented ingrowth at the proximal end and was removed after implantation of a second fully covered stent.

Mechanical examination revealed significantly lower radial expansion force of the new stent as compared to the cylindrical stent.

Conclusions: The new stent design has a low rate of migration. Biomechanical properties may explain this effect.

 
  • References

  • 1 Kahl S, Zimmermann S, Genz I et al. Risk factors for failure of endoscopic stenting of biliary strictures in chronic pancreatitis: a prospective follow-up study. Am J Gastroenterol 2003; 98: 2448-2453
  • 2 Draganov P, Hoffman B, Marsh W et al. Long-term outcome in patients with benign biliary strictures treated endoscopically with multiple stents. Gastrointest Endosc 2002; 55: 680-686
  • 3 Dumonceau JM, Delhaye M, Tringali A et al. Endoscopic treatment of chronic pancreatitis: European Society of Gastrointestinal Endoscopy (ESGE) Clinical Guideline. Endoscopy 2012; 44: 784-800
  • 4 Garcia-Cano J. Use of fully covered self-expanding metal stents in benign biliary diseases. World J Gastrointest Endosc 2012; 4: 142-147
  • 5 Kao D, Zepeda-Gomez S, Tandon P et al. Managing the post-liver transplantation anastomotic biliary stricture: multiple plastic versus metal stents: a systematic review. Gastrointest Endosc 2013; 77: 679-691
  • 6 Mahajan A, Ho H, Sauer B et al. Temporary placement of fully covered self-expandable metal stents in benign biliary strictures: midterm evaluation (with video). Gastrointest Endosc 2009; 70: 303-309
  • 7 Moon JH, Choi HJ, Koo HC et al. Feasibility of placing a modified fully covered self-expandable metal stent above the papilla to minimize stent-induced bile duct injury in patients with refractory benign biliary strictures (with videos). Gastrointest Endosc 2012; 75: 1080-1085
  • 8 Park do H, Lee SS, Lee TH et al. Anchoring flap versus flared end, fully covered self-expandable metal stents to prevent migration in patients with benign biliary strictures: a multicenter, prospective, comparative pilot study (with videos). Gastrointest Endosc 2011; 73: 64-70
  • 9 Park JK, Moon JH, Choi HJ et al. Anchoring of a fully covered self-expandable metal stent with a 5F double-pigtail plastic stent to prevent migration in the management of benign biliary strictures. Am J Gastroenterol 2011; 106: 1761-1765
  • 10 Tarantino I, Mangiavillano B, Di Mitri R et al. Fully covered self-expandable metallic stents in benign biliary strictures: a multicenter study on efficacy and safety. Endoscopy 2012; 44: 923-927
  • 11 Pausawasadi N, Soontornmanokul T, Rerknimitr R. Role of fully covered self-expandable metal stent for treatment of benign biliary strictures and bile leaks. Korean J Radiol 2012; 13 (Suppl. 01) 67-73
  • 12 Deviere J, Nageshwar Reddy D, Puspok A et al. Successful management of benign biliary strictures with fully covered self-expanding metal stents. Gastroenterology 2014; 147: 385-395; quiz e315
  • 13 Irani S, Baron TH, Akbar A et al. Endoscopic treatment of benign biliary strictures using covered self-expandable metal stents (CSEMS). Dig Dis Sci 2014; 59: 152-160