Back to Journals » Clinical Ophthalmology » Volume 6

Relationship between endothelial cell loss and microcoaxial phacoemulsification parameters in noncomplicated cataract surgery

Authors Mahdy M, Eid, Mohammed, Hafez, Bhatia

Received 18 January 2012

Accepted for publication 1 February 2012

Published 29 March 2012 Volume 2012:6 Pages 503—510

DOI https://doi.org/10.2147/OPTH.S29865

Review by Single anonymous peer review

Peer reviewer comments 3



Mohamed AE Soliman Mahdy1,2, Mohamed Z Eid1, Mahmoud Abdel-Badei Mohammed3, Amr Hafez4,5, Jagdish Bhatia2

1Ophthalmic Department, Al-Hussein University Hospital, Al-Azhar University, Cairo, Egypt; 2Ophthalmic Department, Rustaq Hospital, Rustaq, Sultanate of Oman; 3Research Institute of Ophthalmology, Cairo, Egypt; 4Magrabi Eye and Ear Center, Muscat, Sultanate of Oman; 5Ophthalmic Department, Al-Azhar University Hospital, Assuit, Egypt

Purpose: To assess the relationship between postoperative endothelial cell loss and microcoaxial phaco parameters using Ozil IP (Alcon Laboratories, Inc, Fort Worth, TX) in noncomplicated cataract surgery.
Methods: In this prospective observational study, 120 consecutive cases of cataract patients with different grades of nuclear hardness underwent microcoaxial phacoemulsification through a 2.2-mm clear corneal incision. An Alcon Infinity Vision System with Ozil IP (Alcon Laboratories) was used with an Ozil torsional handpiece and a Kelman-style 45° phacoemulsification tip. Patients underwent preoperative and postoperative central endothelial cell counts.
Results: The study included 120 cases of age-related cataract whose mean age (standard deviation [SD]) was 59.68 years (9.47). There was a highly statistically significant endothelial cell loss (P < 0.001). The endothelial cell loss ranged 11–1149 cells/mm2 with a median (interquartile range) of 386 cells/mm2 (184.5–686 cells/mm2). The percentage of postoperative ECLoss% ranged from 0.48% to 47.8% with a median (interquartile range) of 15.4% (7.2% to 26.8%). A significant positive correlation was found between the ECLoss% and different phaco parameters. The Spearman’s rank-order correlation coefficient values, rho, (ρ) were as follows: CDE (ρ = 0.425), aspiration time (ρ = 0.176), and volume (ρ = 0.278). Also, ECLoss% was significantly correlated with the grade of nuclear opalescence (Kendall’s tau τ = 0.42).
Conclusion: Microcoaxial phacoemulsification was efficient in removing noncomplicated cataracts; however a statistically significant endothelial cell loss was noted, especially with increased nuclear hardness. This endothelial cell loss was mostly related to the increased cumulative dissipated energy (CDE), aspiration time, and volume of balanced salt solution used.

Keywords: cataract surgery, phacoemulsification, Ozil, endothelial cells, cumulative dissipated energy

 

Creative Commons License © 2012 The Author(s). This work is published and licensed by Dove Medical Press Limited. The full terms of this license are available at https://www.dovepress.com/terms.php and incorporate the Creative Commons Attribution - Non Commercial (unported, v3.0) License. By accessing the work you hereby accept the Terms. Non-commercial uses of the work are permitted without any further permission from Dove Medical Press Limited, provided the work is properly attributed. For permission for commercial use of this work, please see paragraphs 4.2 and 5 of our Terms.