Skip to main content

Low-flow CO2removal integrated into a renal-replacement circuit can reduce acidosis and decrease vasopressor requirements

Abstract

Introduction

Lung-protective ventilation in patients with ARDS and multiorgan failure, including renal failure, is often paralleled with a combined respiratory and metabolic acidosis. We assessed the effectiveness of a hollow-fiber gas exchanger integrated into a conventional renal-replacement circuit on CO2 removal, acidosis, and hemodynamics.

Methods

In ten ventilated critically ill patients with ARDS and AKI undergoing renal- and respiratory-replacement therapy, effects of low-flow CO2 removal on respiratory acidosis compensation were tested by using a hollow-fiber gas exchanger added to the renal-replacement circuit. This was an observational study on safety, CO2-removal capacity, effects on pH, ventilator settings, and hemodynamics.

Results

CO2 elimination in the low-flow circuit was safe and was well tolerated by all patients. After 4 hours of treatment, a mean reduction of 17.3 mm Hg (−28.1%) pCO2 was observed, in line with an increase in pH. In hemodynamically instable patients, low-flow CO2 elimination was paralleled by hemodynamic improvement, with an average reduction of vasopressors of 65% in five of six catecholamine-dependent patients during the first 24 hours.

Conclusions

Because no further catheters are needed, besides those for renal replacement, the implementation of a hollow-fiber gas exchanger in a renal circuit could be an attractive therapeutic tool with only a little additional trauma for patients with mild to moderate ARDS undergoing invasive ventilation with concomitant respiratory acidosis, as long as no severe oxygenation defects indicate ECMO therapy.

Introduction

In modern intensive care medicine, patients with multiorgan failure frequently need respiratory ventilation and renal-replacement therapy to bridge organ dysfunction. Ventilation itself, in particular with the use of high tidal volumes and high airway pressures, has been shown to be deleterious for patient outcomes [1, 2], and thus protective ventilation strategies, including lower tidal volumes, have been implemented into clinical practice [1, 3]. Although these ventilation specifications often lead to respiratory acidosis, the concept of permissive hypercapnia and concomitant acidosis is presently widely accepted, albeit still controversially discussed in the scientific community. Whereas evidence has been provided for immunologic, redox, and vasoactive protective effects, acidosis has also been associated with higher hemodynamic instability [47]. Extracorporeal membrane oxygenation (ECMO) [2, 810] and pumpless CO2-removal systems (PECLA) [11, 12] are increasingly used to support lung-protective ventilation strategies and to improve CO2 removal and respiratory acidosis. Importantly, in the recent Xtravent study by Bein et al.[13], pumpless CO2 removal enabled efficient low-tidal ventilation (about 3 ml/kg PBW) without severe acidosis, which was also associated with more ventilator-free days for patients having a severe oxygenation deficit (PaO2/FIO2 <150). However, in all of these therapies, a higher extracorporeal blood-flow circuit with additional risk and trauma by using special cannulas with big diameters to the patient is mandatory.

About 35% to 60% of the patients undergoing respiratory therapies in multiorgan failure also need renal-replacement therapies (RRTs) [14, 15]. Because, in these patients with concomitant renal failure, extracorporeal blood circuits have necessarily already been established for renal-replacement therapy, we wondered whether addition of a hollow-fiber gas exchanger to the low-flow blood circuit could support lung-protective strategies by improving respiratory acidosis. Although renal-replacement circuits allow only a blood flow of 300 to 500 ml/min, partial elimination of CO2 appears to be feasible [16]. Arterial blood with a pCO2 of 40 mm Hg (5.3 kPa) contains approximately 500 ml CO2/L (pH 7.4). In sheep experiments, Young et al.[17] achieved 130 to 180 ml CO2 elimination (500 ml/L pCO2 in blood at 40 mm Hg (5.3 kPa) calculated) by using blood-flow rates of about 500 ml/min, combining a hollow-fiber gas exchanger and a hemofiltration device. Livigni et al.[18] also tested effects of low-flow CO2 elimination in sheep by using a veno-venous pump-driven bypass. They found an average CO2 reduction of hypercapnic ventilated sheep of 17% to 22%. Batchinsky et al.[19] were able to achieve normocapnia in hypercapnic ventilated pigs by using a veno-venous pump-driven system (400 to 600 ml/min blood flow), including a gas exchanger. Altogether, experimental evidence suggests, therefore, that a significant amount of basal CO2 production can be eliminated in low-flow veno-venous systems. First experiences in critically ill patients applying CO2 removal with low-flow veno-venous systems were gained by using a specialized device with a hollow-fiber gas exchanger adapted to low blood flows (about 350 ml/min; DecapSmart”). First, case reports described successful application of this system in single patients [2023]. Eventually, a clinical study using the DecapSmart system was able to demonstrate effects of low-flow CO2 removal in critically ill patients. Here in 10 patients being ventilated with a plateau pressure between 28 and 30 cm H2O and having respiratory acidosis, additional low-flow CO2 removal reduced pCO2 from 73.6 to 50.4 mm Hg and increased pH from 7.20 to 7.32 in 60 to 90 minutes [24]. However, the DecapSmart system still needs a specialized device and cannulation, which needs vascular access side to side with cannulas needed for renal-replacement therapy in severe AKI. We thus wondered whether implementation of a hollow-fiber gas exchanger into the renal circuit could combine low-flow CO2 removal and renal-replacement therapy by using one system and blood access and tested for ventilatory and hemodynamic effects in 10 severely ill patients with combined respiratory and renal failure.

Materials and methods

CRRT circuit

We used a CVVHD device (bm11/14; Edwards-Lifescience, Irvine, CA, USA) with a standard setup and adjustment for continuous venovenous hemodialysis. A high-flux polysulfone capillary hemofilter with a membrane surface area of 1.4 m2 (Polyflux 140 H; Gambro, Hechingen, Germany ) was used.

For decarboxylation, a small standard hollow-fiber gas exchanger (D902 Liliput 2 ECMO; Sorin Group Milan, Italy) was applied. This gas exchanger has a surface area of 0.67 m2 and is intended for extracorporeal circuits with a maximum blood flow of 2,300 ml/min. According to the manufacturer’s description, the filter is coated with phosphorylcholine, which should form a phospholipid-like structure and reduce coagulation. The gas exchanger was integrated into the continuous hemodialysis system after the dialysis filter. For the connection, we used standard tubes with a Luer-lock system. The gas exchanger was attached to the hemodialysis machine with a conventional clamp (Figure 1). Venous and arterial pressures were monitored continuously. Gas flow through the gas exchanger was set to 4 L/min (blood flow, <300 ml/min) or 4 to 6 L/min (blood flow, >300 ml/min) with a FiO2 of 0.21. Only in cases in which oxygenation by the ventilator was critical, the FiO2 was varied (0.5 to 1.0 FiO2), in the hope of counteracting systemic hypoxia. For reasons of simplicity, we called this setting “lung-assisting renal replacement system” or “LARRS.”

Figure 1
figure 1

Scheme of the renal circuit with implementation of a hollow-filter gas exchanger.

The LARRS system was tested beforehand in “dummy” experiments, yielding no detectable changes in venous, arterial, and transmembrane pressures and confirming the full function of alarms, bubble catcher, and emergency shutter. In particular, flushing the system with air by artificial occlusion of all three ventilation tubes of the oxygenator led to prompt emergency standby of the CVVHD circuit, according to security standards in renal-replacement therapy devices. Also when the system was subsequently applied to patients, the pressure control of the hemodialysis did not show any differences in the presence of the gas exchanger in comparison to conventional renal-replacement therapy.

Blood flow was primarily adjusted to 400 ml/min, but could be adapted to individual needs and circumstances (300 to 500 ml/min). A 13.5-French double-lumen catheter was placed into the jugular vein. In two patients, we used a second 13.5-French double-lumen catheter in the femoral vein to allow higher blood flows. In this case, the two lumina of one catheter were linked by a y-adapter to form a single blood line. This allowed increasing the blood flow about 30% to 40% in these two patients. The dialysate contained Na, 140 mM; K, 4.0 M; Ca, 1.5 mM; Mg, 0.5 mM; Cl, 113 mM; HCO3, 35.0 mM; and D-Glucose, 9.0 g/L. Anticoagulation was performed with systemic heparin, and doses were prescribed targeting a PTT of 60 seconds or an activated clotting time (ACT), which was measured in bedside assays of 160 to 200 seconds. To prevent cooling of the patient, we used a tube heating system (Fresenius, Germany), which was set to 37°C.

Patient inclusion

Patients were treated with the hollow-fiber gas exchanger in the renal circuit according to the individual decision of the treating physician, based on the patient’s characteristics and needs. Inclusion criteria were primarily the need for renal-replacement and mechanical-ventilation therapy with concomitant hypercapnic respiratory acidosis (pH <7.25). The treatment protocol was approved by the local Ethics Committee of the University of Erlangen-Nuremberg, Germany. In all cases, written informed consent was obtained from a legal guardian of the patient before application of the gas-exchange filter was started.

All patients had a central venous catheter in addition to the double-lumen catheters used for RRT, a urinary catheter (unless they were anuric), and an invasive blood-pressure measurement. Heart rate, blood pressure, SaO2, and temperature were monitored continuously; diuresis was measured hourly, and arterial blood gases were measured in variable intervals during the whole stay at the intensive care unit. Norepinephrine infusions were applied in parallel to fluid administration to maintain a mean arterial pressure of 65 mm Hg.

Statistics

Because of the small number of patients, most results were presented for each case in absolute values or by calculating differences between baseline and 24 hours after commencing treatment. Means are expressed as mean ± SD of the mean.

Results

Patient characteristics

We treated 10 patients with the hollow-filter gas exchanger (LARRS) between November 2009 and January 2011. All patients were already undergoing CVVHD because of acute renal failure and oliguria when the CO2 hollow filter was applied. At inclusion, the arterial pH varied from 7.07 to 7.24; patient 7 was included to prevent progressive respiratory failure and exhaustion due to a still-compensated respiratory acidosis (HCO3, 31 mM; pCO2, 55 mm Hg; pH 7.37). The patients’ baseline characteristics are summarized in Table 1. Eight patients had community-acquired pneumonia; the other two patients had acute infectious exacerbation of COPD. Two patients (1 and 2) fulfilled ARDS criteria, but had contraindications against ECMO therapy. One patient (5) with severe ARDS was treated briefly for bridging until ECMO therapy was installed. Although this patient had preexisting high bicarbonate values (37 mm Hg) because of long-term adaptation to his COPD, this patient developed progressive respiratory acidosis (pH 7.34; pCO2, 88 mm Hg) owing to his severe respiratory failure and increasing difficulties in ventilation support. The mean APACHE II score was 29.6 (APACHE II, 22 to 35), with a mean predicted mortality of 72% (40% to 82%).

Table 1 Baseline patients’ characteristics and outcomes of the 10 patients treated with the hollow-fiber gas-exchange device included in the renal-replacement circuit

CO2elimination in the CRRT circuit

Treatment modalities are summarized in Table 2. Application of the hollow-fiber gas-exchanger CVVHD system was well tolerated in all cases, and no change of function of the CVVHD device and the renal-replacement filter was detectable or necessary. Blood-flow rates from 250 ml to 500 ml could be achieved in all patients (mean, 378 ml/min at 24 hours). In particular, intraluminal pressures in the system were not altered. All patients tolerated the intervention well, and no complications occurred during the therapy. Mean heparin doses were 815 (t = 0 hours) and 1,116 IU/h (t = 24 hours), resulting in a mean PTT of 43 (t = 0 hours) and 72 (t = 24 hours) seconds. Mean activated clotting time was 177 (t = 0 hours) and 213 seconds (t = 24 hours) in bedside tests (see Additional file 1: Table S1). Two episodes of clotting were observed. In one case, the renal filter clotted after 23 hours of use and had to be exchanged. In a second case, the hollow-fiber gas exchanger clotted, which led to a rapid decrease of the arterial pH and increase of pCO2 in the extracorporeal system and activated pressure alarms of the dialysis machine. In this case, the hollow-fiber gas exchanger was exchanged immediately, which quickly restored CO2 removal and pH compensation.

Table 2 Treatment modalities for low-flow CO 2 removal integrated into the RRT circuit

Overall, in-device blood gas analysis directly after the hollow-fiber gas exchanger showed a mean reduction of extracorporeal pCO2 of 39 mm Hg (63 mm Hg before gas-exchanger to 24 mm Hg after gas-exchanger filter; see Additional file 2: Figure S1), resulting in a mean increase of pH of 0.31 (pH 7.28 before filter and 7.59 after filter, mean blood flow of 377 ml/min). The pO2 in the extracorporeal system increased from 41 mm Hg (before filter) to 122 mm Hg (after filter), but no change occurred in the patients’ systemic pO2. (Additional file 2: Figure S1).

Blood-gas changes during the course of therapy

Implementation of the hollow-fiber gas exchanger reduced the average systemic arterial pCO2 by 17.3 mm Hg or 28.1% in about 4 hours (Table 3). The average pCO2 could then be kept constant during the next 24 hours. The pH concomitantly increased by 0.12 (0.04 to 0.19) in the first 4 hours, remaining approximately constant over the next 24 hours. No change was seen in arterial bicarbonate concentrations, presumably due to continuous dialysis against bicarbonate-containing dialysate. Despite an FiO2 of 1.0 in the gas flow to the gas-exchange device and a concomitant increase in the extracorporeal pO2 in the first two patients, no change occurred in their arterial pO2, which is consistent with the dominant role of pulmonary gas exchange for arterial pO2. The absence of a measurable effect on oxygenation can most likely be attributed to the low blood flow and thus to the low amount of oxygen provided for systemic circulation. We therefore subsequently reduced the FiO2 at the membrane oxygenator to 0.21 to avoid potential side effects or counterregulatory vascular effects in the pulmonary circulation arising from possibly increased oxygen tensions inside the pulmonary artery and pulmonary capillary bed. Despite the improvements in CO2 elimination, ventilator settings remained unchanged or were slightly deescalated. In particular, at t = 24 hours, mean Pmax could be reduced from 32.50 to 28.67 mm Hg at t = 24 hours. Mean tidal volumes were slightly decreased from 8.41 to 7.34 ml PBW, eventually still not achieving 6 ml/kg PBW (Table 4).

Table 3 Changes in systemic arterial pH, pCO 2 , bicarbonate (HCO 3 ) from the beginning ( t= 0 hours), 4 hours ( t= 4 hours), and 24 hours ( t= 24 hours) after starting low-flow CO 2 removal
Table 4 Ventilator settings from the beginning ( t= 0 hours), 4 hours ( t= 4 hours) and 24 hours ( t= 24 hours) after starting low-flow CO 2 removal

Hemodynamic stability

In parallel to pH correction, a marked stabilization of hemodynamics was observed (Figure 2). When the hollow-fiber gas-exchanger CVVHD system was started, nine of 10 patients received norepinephrine therapy, five of these in doses between 0.6 and 5.0 mg/h (0.13 to 0.93 μg/kg/min). In four of these five patients with higher doses, a marked hemodynamic stabilization with decreased norepinephrine doses (average 65% reduction) could be achieved during the therapy (Table 5). In the remaining hemodynamically unstable patient (number 1), acidosis could not be corrected, and no hemodynamic stabilization occurred. In terms of volume control, patients were either ultrafiltrated because of volume overload in lung failure in the first 24 hours (four patients, -645 to −3,345 ml/24 hours) or received volume replacement as part of hypertensive shock therapy (six patients, 180 to 2,945 ml/24 hours). No obvious difference was seen in the response to CO2 elimination between patients in whom either negative or positive fluid balance was achieved. Exemplary time courses of pH, pCO2, and norepinephrine dose in two patients who were successfully treated with the low-flow RRT system are shown in Figure 3.

Figure 2
figure 2

Clinical effects of low-flow CO 2 removal 4 and 24 hours after commencement of treatment. Graphs illustrate mean changes of values for pH (A), pCO2(B), pO2(C), tidal volume on respirator (D), norepinephrine dose at mg/h (E), and average changes in Pmax on the ventilator (F).

Figure 3
figure 3

Exemplary time courses for clinical effects of low-flow CO 2 removal in two patients. Patient 10 had acute pneumonia (A, B, C); patient 6 had cardiomyopathy and acute pneumonia (D, E, F) for pH, pCO2, and norepinephrine dose, respectively.

Table 5 Hemodynamic parameters from the beginning ( t= 0 h), 4 hours ( t= 4 h), and 24 hours ( t= 24 h) after starting low-flow CO 2 removal

Patients’ outcome

No serious adverse events could be attributed to the hollow-fiber gas exchanger or the RRT circuit during treatment of the patients. Seven of 10 patients were successfully weaned from the low-flow CO2 removal system, improved their pulmonary function, and recovered from critical illness (Table 1).

In the two patients (patients 1 and 2), who had contraindications for ECMO therapy, low-flow CO2 removal was applied into the preexisting CVVHD devices as an additional limited therapy option in very severe multiple organ failure: both patients had H1N1 infection with pulmonary ARDS and very low oxygenation indices (95/91). Patient 1 had also a severe liver failure secondary to preexisting alcoholic liver cirrhosis Child C. He died at day 3 after study inclusion. Patient 2 had Hodgkin disease and was undergoing chemotherapy. He had acquired H1N1 and concomitant Aspergillus fumigatus pneumonia. We eventually limited intensive care support because of the overall poor prognosis and in line with the patient’s provision. He died at day 4 after commencing intensive care therapy. Patient 5, with ARDS due to pneumonia, was bridged to ECMO therapy for only 2.5 hours, before ECMO cannulas were applied and an ECMO device was available. He died 12 days later of multiorgan failure despite continued ECMO therapy.

Discussion

For the first time we here report the use of a low-flow hollow-fiber gas exchanger implemented in a CRRT circuit in a small series of critically ill patients. The simple device proved to be efficient in terms of CO2 elimination, was well tolerated, and did not lead to adverse events. Concomitant renal-replacement therapy was in no way compromised, and alarm functions of the CRRT system ensured safety control for the gas-exchange device.

The concept of permissive hypercapnia has been developed to reduce baro- and volutrauma during ventilation. Increased CO2 levels have been shown to be associated with some potentially beneficial (for example, antiinflammatory effects), but the resulting acidosis also induces hemodynamic instability and potential cellular adverse effects [47]. In our case series in eight of 10 patients, application of the gas-exchange filter led to a rapid, partial, or complete correction of the pH and a significant reduction of the pCO2 within 4 hours. Ranges of CO2 reduction and pH correction were overall comparable to results obtained in the previous study of Terragni et al. [24] using a standalone low-flow CO2-removal system. In five of six hemodynamically unstable patients requiring higher doses of norepinephrine (>0.5 mg/h/>0.1 μg/kg/min), pH correction was in line with a marked reduction of vasopressor needs and an improved hemodynamic stability. Norepinephrine doses could be reduced to approximately one half after 6 to 8 hours on average; in three patients, vasopressors could even be stopped. Hemodynamic stability correlated with pH correction by enhanced CO2 elimination.

In the CRRT circuit used, blood flow was limited to 500 ml/min at maximum, provided that catheter flow was optimal. We here used a high ratio of blood to gas flow to achieve maximum CO2 elimination, although lower blood/gas-flow ratios would be potentially effective as well. Achieved CO2 removal allowed correction of acidosis, but we did not find a significant contribution to the systemic oxygenation of the patients even when higher FiO2 together with high gas/blood-flow ratios were applied. Thus, additional systemic oxygenation cannot be achieved with this low-flow device and is reserved for extracorporeal lung replacement therapies with higher flow rates, as in ECMO or PECLA therapy.

As it was a pilot, nonrandomized proof-of-concept study, our study has several inherent limitations. First, implementation of the device was one of multiple interventions in the severely ill patients under investigation. Treatment included volume administration, antibiotic treatment, and further co-medication to treat the underlying disease, which was sepsis in eight of 10 cases. Treatment of acidosis was not confined to CO2 elimination by the respirator and the hollow-fiber gas exchanger, but also included bicarbonate filtrate substitution and proton dialysis during the course of renal-replacement therapy. However, the time course of blood pH and pCO2 in individual patients with immediate responses after the implementation of the hollow-fiber gas exchanger clearly argue for a causal impact of the device. Moreover, metabolic acidosis was already balanced by bicarbonate dialysis, when low-flow CO2 removal was started, and bicarbonate levels remained more or less stable during renal-replacement therapy and therefore most likely did not contribute to the observed effects. Second, we did not apply a standardized ventilation protocol, which would have allowed testing for effects of the intervention on ventilatory requirements or achieving low tidal volumes of 6 ml/kg PBW or less. However, in our clinical setting, the hollow-fiber gas exchanger in a low-flow circuit CO2 at least compensated for respiratory acidosis and avoided further increases in ventilation settings or even led to slight reductions of peak pressure despite ongoing respiratory failure. Statistically, the overall survival in this small series was much higher than predicted by the calculated APACHE II scores of the patients, but a sample size of 10 patients certainly does not allow drawing any conclusions in respect to outcomes.

Despite this principal limitation of the hollow-fiber gas exchanger inserted into an RRT circuit, the system may have significant advantages in settings where the control of hypercapnia is desirable, including less-severe ARDS, when ECMO therapy is either still not indicated or simply not available. The advantages of applying a hollow-fiber gas exchanger in a CRRT circuit include its simplicity and its potential applicability in nonspecialized centers, which are experienced in renal-replacement therapy, as well as the fact that no additional catheter placements are needed. In contrast to many ECMO devices, available RRT machines are more secure and are equipped with distinct alarm systems, which allow emergency shut-off due to air or clots in the blood circuits. Regional anticoagulation with citrate was not performed in conjunction with the described setting, because of the necessity for increased blood flows, but regional anticoagulation protocols could possibly be developed, because a hemofilter allowing clearance of calcium-citrate complexes is included in the system, which is not the case in systems with isolated hollow-fiber gas exchangers.

Conclusions

Implementation of a hollow-fiber gas exchanger in a low-flow CRRT circuit was feasible and safe and let to a significant CO2 removal and rapid correction of arterial pH in critically ill patients with acute renal and respiratory failure, with a positive impact on hemodynamic stability. Integration of a hollow-fiber gas exchanger could thus be potentially an additive tool in the armamentarium of treatment modalities in patients with multiorgan failure. To this end, additional, larger, and controlled studies are certainly needed to assess the impact of low-flow CO2 removal on ventilator settings and patient prognosis.

Key messages

  • Implementation of a hollow-fiber gas exchanger in a low-flow CRRT circuit was feasible and safe

  • Low-flow CO2 removal in a CRRT circuit significantly removed CO2 and allowed rapid correction of arterial pH in critically ill patients with acute renal and respiratory failure

  • Correction of arterial pH contributed to hemodynamic stabilization of the patient

  • Low-flow CO2 did not significantly contribute to systemic oxygenation

  • Low-flow CO2 removal complements but did not substitute ECMO or PECLA therapy, when ECMO/PECLA was indicated

Abbreviations

ACT:

Activated clotting time

AKI:

Acute kidney injury

ALI:

Acute lung injury

ARDS:

Acute respiratory distress syndrome

CRRT:

Continuous renal-replacement therapy

CVVHD:

Continuous veno-venous hemodialysis

ECMO:

Extracorporeal membrane oxygenation

LARRS:

Lung-assisting renal-replacement system

PBW:

Predicted body weight

PECLA:

Pumpless extracorporeal lung assist

RRT:

Renal-replacement therapy.

References

  1. Malhotra A: Low-tidal-volume ventilation in the acute respiratory distress syndrome. N Engl J Med. 2007, 357: 1113-1120. 10.1056/NEJMct074213.

    Article  PubMed  CAS  PubMed Central  Google Scholar 

  2. Walkey AJ, Summer R, Ho V, Alkana P: Acute respiratory distress syndrome: epidemiology and management approaches. J Clin Epidemiol. 2012, 4: 159-169.

    Article  Google Scholar 

  3. Amato MB, Barbas CS, Medeiros DM, Magaldi RB, Schettino GP, Lorenzi-Filho G, Kairalla RA, Deheinzelin D, Munoz C, Oliveira R, Takagaki TY, Carvalho CR: Effect of a protective-ventilation strategy on mortality in the acute respiratory distress syndrome. N Engl J Med. 1998, 338 (6): 347-354. 10.1056/NEJM199802053380602.

    Article  PubMed  CAS  Google Scholar 

  4. O’Croinin D, Ni Chonghaile M, Higgins B, Laffey JG: Bench-to-bedside review: permissive hypercapnia. Crit Care. 2005, 9: 51-59.

    Article  PubMed  Google Scholar 

  5. Curley G, Contreras MM, Nichol AD, Higgins BD, Laffey JG: Hypercapnia and acidosis in sepsis: a double-edged sword?. Anesthesiology. 2010, 112: 462-472. 10.1097/ALN.0b013e3181ca361f.

    Article  PubMed  Google Scholar 

  6. Ijland MM, Heunks LM, van der Hoeven JG: Bench-to-bedside review: hypercapnic acidosis in lung injury–from “permissive” to”therapeutic. Crit Care. 2010, 14: 237-

    Article  PubMed  Google Scholar 

  7. Ismaiel NM, Henzler D: Effects of hypercapnia and hypercapnic acidosis on attenuation of ventilator-associated lung injury. Minerva Anestesiol. 2011, 77: 723-733.

    PubMed  CAS  Google Scholar 

  8. Sidebotham D, McGeorge A, McGuinness S, Edwards M, Willcox T, Beca J: Extracorporeal membrane oxygenation for treating severe cardiac and respiratory failure in adults: part 2, technical considerations. J Cardiothorac Vasc Anesth. 2010, 24: 164-172. 10.1053/j.jvca.2009.08.002.

    Article  PubMed  Google Scholar 

  9. Allen S, Holena D, McCunn M, Kohl B, Sarani B: A review of the fundamental principles and evidence base in the use of extracorporeal membrane oxygenation (ECMO) in critically ill adult patients. J Intensive Care Med. 2011, 26: 13-26. 10.1177/0885066610384061.

    Article  PubMed  Google Scholar 

  10. Peek GJ, Mugford M, Tiruvoipati R, Wilson A, Allen E, Thalanany MM, Hibbert CL, Truesdale A, Clemens F, Cooper N, et al: Efficacy and economic assessment of conventional ventilatory support versus extracorporeal membrane oxygenation for severe adult respiratory failure (CESAR): a multicentre randomised controlled trial. Lancet. 2009, 374: 1351-1363. 10.1016/S0140-6736(09)61069-2.

    Article  PubMed  Google Scholar 

  11. Bein T, Zimmermann M, Hergeth K, Ramming M, Rupprecht L, Schlitt HJ, Slutsky AS: Pumpless extracorporeal removal of carbon dioxide combined with ventilation using low tidal volume and high positive end-expiratory pressure in a patient with severe acute respiratory distress syndrome. Anaesthesia. 2009, 64: 195-198. 10.1111/j.1365-2044.2008.05735.x.

    Article  PubMed  CAS  Google Scholar 

  12. Zimmermann M, Bein T, Arlt M, Philipp A, Rupprecht L, Mueller T, Lubnow M, Graf BM, Schlitt HJ: Pumpless extracorporeal interventional lung assist in patients with acute respiratory distress syndrome: a prospective pilot study. Crit Care. 2009, 13: R10-10.1186/cc7703.

    Article  PubMed  PubMed Central  Google Scholar 

  13. Bein T, Weber-Carstens S, Goldmann A, Muller T, Staudinger T, Brederlau J, Muellenbach R, Dembinski R, Graf BM, Wewalka M, et al: Lower tidal volume strategy (approximately 3 ml/kg) combined with extracorporeal CO2 removal versus”conventional” protective ventilation (6 ml/kg) in severe ARDS: the prospective randomized Xtravent study. Intensive Care Med. 2013, 39: 847-856. 10.1007/s00134-012-2787-6.

    Article  PubMed  PubMed Central  Google Scholar 

  14. Esteban A, Alia I, Gordo F, de Pablo R, Suarez J, Gonzalez G, Blanco J: Prospective randomized trial comparing pressure-controlled ventilation and volume-controlled ventilation in ARDS, for the Spanish Lung Failure Collaborative Group. Chest. 2000, 117: 1690-1696. 10.1378/chest.117.6.1690.

    Article  PubMed  CAS  Google Scholar 

  15. Liu KD, Matthay MA: Advances in critical care for the nephrologist: acute lung injury/ARDS. Clin J Am Soc Nephrol. 2008, 3: 578-586. 10.2215/CJN.01630407.

    Article  PubMed  CAS  Google Scholar 

  16. Gattinoni L, Kolobow T, Tomlinson T, Iapichino G, Samaja M, White D, Pierce J: Low-frequency positive pressure ventilation with extracorporeal carbon dioxide removal (LFPPV-ECCO2R): an experimental study. Anesth Analg. 1978, 57: 470-477.

    Article  PubMed  CAS  Google Scholar 

  17. Young JD, Dorrington KL, Blake GJ, Ryder WA: Femoral arteriovenous extracorporeal carbon dioxide elimination using low blood flow. Crit Care Med. 1992, 20: 805-809. 10.1097/00003246-199206000-00018.

    Article  PubMed  CAS  Google Scholar 

  18. Livigni S, Maio M, Ferretti E, Longobardo A, Potenza R, Rivalta L, Selvaggi P, Vergano M, Bertolini G: Efficacy and safety of a low-flow veno-venous carbon dioxide removal device: results of an experimental study in adult sheep. Crit Care. 2006, 10: R151-10.1186/cc5082.

    Article  PubMed  PubMed Central  Google Scholar 

  19. Batchinsky AI, Jordan BS, Regn D, Necsoiu C, Federspiel WJ, Morris MJ, Cancio LC: Respiratory dialysis: reduction in dependence on mechanical ventilation by venovenous extracorporeal CO2 removal. Crit Care Med. 2011, 39: 1382-1387. 10.1097/CCM.0b013e31820eda45.

    Article  PubMed  Google Scholar 

  20. Ruberto F, Pugliese F, D’Alio A, Perrella S, D’Auria B, Ianni S, Anile M, Venuta F, Coloni GF, Pietropaoli P: Extracorporeal removal CO2 using a venovenous, low-flow system (Decapsmart) in a lung transplanted patient: a case report. Transplant Proc. 2009, 41: 1412-1414. 10.1016/j.transproceed.2009.03.048.

    Article  PubMed  CAS  Google Scholar 

  21. Iacovazzi M, Oreste N, Sardelli P, Barrettara B, Grasso S: Extracorporeal carbon dioxide removal for additional pulmonary resection after pneumonectomy. Minerva Anestesiol. 2012, 78: 381-384.

    PubMed  CAS  Google Scholar 

  22. Moscatelli A, Ottonello G, Nahum L, Lampugnani E, Puncuh F, Simonini A, Tumolo M, Tuo P: Noninvasive ventilation and low-flow veno-venous extracorporeal carbon dioxide removal as a bridge to lung transplantation in a child with refractory hypercapnic respiratory failure due to bronchiolitis obliterans. Pediatr Crit Care Med. 2010, 11: e8-e12. 10.1097/PCC.0b013e3181b0123b.

    Article  PubMed  Google Scholar 

  23. Gramaticopolo S, Chronopoulos A, Piccinni P, Nalesso F, Brendolan A, Zanella M, Cruz DN, Ronco C: Extracorporeal CO2 removal: a way to achieve ultraprotective mechanical ventilation and lung support: the missing piece of multiple organ support therapy. Contrib Nephrol. 2010, 165: 174-184.

    Article  PubMed  Google Scholar 

  24. Terragni PP, Del Sorbo L, Mascia L, Urbino R, Martin EL, Birocco A, Faggiano C, Quintel M, Gattinoni L, Ranieri VM: Tidal volume lower than 6 ml/kg enhances lung protection: role of extracorporeal carbon dioxide removal. Anesthesiology. 2009, 111: 826-835. 10.1097/ALN.0b013e3181b764d2.

    Article  PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Carsten Willam.

Additional information

Competing interests

Sorin (Milan, Italy) provided the hollow-fiber gas exchangers without cost. The authors declare that they have no competing interests.

Authors’ contributions

CF participated in the design of the study, carried out the data acquisition and analysis, and participated in drafting the manuscript. JS primarily established the LARRS system and controlled the technical application. SJ and KUE participated in the design of the study and participated in drafting the manuscript. CW conceived of and designed the study, carried out the coordination, and drafted the manuscript. All authors read and approved the final manuscript.

Electronic supplementary material

13054_2013_1896_MOESM1_ESM.doc

Additional file 1: Table S1: Anticoagulation therapy and clotting values after starting low-flow CO2 removal. n.a., not applicable. (DOC 48 KB)

13054_2013_1896_MOESM2_ESM.tiff

Additional file 2: Figure S1: In device pre- and postfilter pH, pCO2 and pO2 blood-gas measurements values. (TIFF 205 KB)

Authors’ original submitted files for images

Rights and permissions

Reprints and permissions

About this article

Cite this article

Forster, C., Schriewer, J., John, S. et al. Low-flow CO2removal integrated into a renal-replacement circuit can reduce acidosis and decrease vasopressor requirements. Crit Care 17, R154 (2013). https://doi.org/10.1186/cc12833

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/cc12833

Keywords