Introduction

Glaucoma affects approximately 67 million people making it the most common cause of irreversible blindness worldwide.1 Although constituting only about 26% of all glaucoma, primary angle closure glaucoma (PACG) is an aggressive and visually destructive disease,2, 3 with estimations that it blinds five times more people than primary open-angle glaucoma4 making it an important public health problem. The prevalence of PACG varies among different ethnic groups and is a major form of glaucoma in the populous nations of China5 and India.6 To effectively prevent PACG by the use of prophylactic laser iridotomy, it is necessary to identify people with anatomically narrow angles.

Anterior segment optical coherence tomography (ASOCT) offers a non-contact method for detecting eyes at risk for angle closure.7, 8, 9 Several recent studies have used ASOCT to visualize anterior chamber angle (ACA) structures7, 10, 11, 12, 13, 14 and demonstrated it to be accurate for the detection of closed or occludable angles, with gonioscopy used as the reference.7, 15, 16 Spectral domain ASOCT (SD-ASOCT) technology is faster and more efficient than the conventional time domain ASOCT (TD-ASOCT): by detecting signals from the entire depth range in parallel, rather than serially, it achieves higher speed without losing the signal-to-noise ratio.17, 18

Scheimpflug imaging also offers non-contact imaging of the anterior segment. Although Scheimpflug photography cannot fully visualize the entire angle,19 the software available on the Pentacam Scheimpflug camera (Oculus, Wetzlar, Germany) provides extrapolated measurements of the anterior chamber volume (ACV), which have shown promise in screening for narrow angles.20

The aim of this investigation was to compare the diagnostic performance of Scheimpflug imaging and SD-ASOCT in detecting narrow angles and to evaluate the correlations among the quantitative parameters obtained from these two devices.

Materials and methods

The study was approved by the hospital's Institutional Review Board and adhered to the tenets of the Declaration of Helsinki. Informed consent was obtained from all participants. Patients aged ≥40 years were recruited from the comprehensive ophthalmology clinic at Grewal Eye Institute. Patients were excluded if they had a history of glaucoma, intraocular surgery, laser treatment, penetrating trauma, and corneal disorders, such as corneal endothelial dystrophy, corneal opacity, pterygium, or abnormalities that precluded SD-ASOCT or Scheimpflug imaging. All participants underwent best-corrected visual acuity (BCVA) measurement using a logarithm of minimum-angle-of-resolution chart (logMAR chart, Lighthouse Inc., Long Island, NY, USA), auto-refraction (Topcon RM-8000B, Topcon Corporation, Tokyo, Japan), slit-lamp examination (Model BQ 900, Haag-Streit, Bern, Switzerland), stereoscopic optic-disc examination with a 78-diopter lens (Volk Optical Inc., Mentor, OH, USA), and intraocular pressure measurement (IOP) using Goldmann applanation tonometry (Haag-Streit, Koniz, Switzerland).

Patients underwent Scheimpflug imaging, SD-ASOCT imaging, and gonioscopy in that order. Scheimpflug and SD-ASOCT imaging were performed in a dark room (1 lux using a digital light meter (Extech 401027 Pocket Digital Light Meter, Extech Instruments, Waltham, MA, USA)) after allowing dark adaptation for 30 s, without the use of any mydriatics, and with the subjects sitting in front of the instrument with their face in an upright position, by a single examiner who was masked to the gonioscopy results. The intensity of light during each examination was measured to standardize the illumination conditions. Manufacturer-recommended quality criteria for image acquisition were used as guidelines for acceptance of scan images.

We certify that all applicable institutional and governmental regulations concerning the ethical use of human volunteers were followed during this research.

Scheimpflug imaging

The Pentacam 50 scan image acquisition mode was used, which captures 50 slit images of the anterior segment in 2 s using a rotating Scheimpflug camera. Only images with a Quality Score >95 were included. The Pentacam software (Software version 1.11, Oculus) configures a three-dimensional model of the anterior segment with the obtained data, and the central anterior chamber depth along the optical axis (ACD, mm), ACA (degrees), and ACV (μl) are computed. The ACV is calculated using an integral calculus, which considers the anterior chamber as a solid bounded by the posterior surface of the cornea (12 mm diameter around the corneal apex), the iris, and the lens.21 Pentacam ACA measurements were not used in this study, as their reliability in eyes with narrow angles has been questioned due to the inability of Scheimpflug imaging to visualize the most peripheral part of the iris and base of the ACA.20

Spectral domain anterior segment optical coherence tomography

The RTVue 100 (Optovue Inc., Fremont, CA, USA, software version 4.0) SD-OCT system was used for anterior segment imaging with an auxiliary lens attachment, the corneal adaptor module (CAM). CAM includes adaptor lenses to produce a telecentric scan geometry for corneal imaging and scan patterns, along with the image processing software. The spectrometer is fitted with a high-speed line camera that captures 26 000 axial (A) scans per second using an 830-nm wavelength light-source providing an axial resolution of 5 μm and a transverse scan resolution of 15 μm.

For this study, anterior segment morphology was assessed with the corneal adaptor module long (CAM-L), using the angle scan protocol, which captured 1 × 1024 A-scans in 0.04 s. The patient's fixation was directed to the side of the instrument, using an external fixation light, so that the irido-corneal angle was centered in the instrument's field of view. The external fixation target was approximately 20 cm away from the patient's eye. The working distance between CAM-L and the cornea was approximately 13 mm. Two scans of the angle at 3 o’clock and 9 o’clock position were obtained, representing the temporal and nasal quadrants of each eye. In order to obtain images with the corneo-scleral surface perpendicular to the scanning beam, the software provides guide marks on the image acquisition screen. The superior and inferior quadrants were not scanned to avoid potential eyelid obstruction or anterior segment distortion from eyelid manipulation. To achieve a non-accommodated state, the power setting was adjusted according to the patient's refraction. Only images with Scan-Score Index >45 were included. For quantitative analysis, an ophthalmologist (SPSG) marked out the scleral-spur location in the nasal and temporal quadrants, following which the software computed the quantitative analysis. The scleral spur was determined as the point at which there was a change in curvature of the inner surface of the angle wall.22 Parameters measured were: (a) angle opening distance at 500 μm from the scleral spur (AOD500), defined by Pavlin et al23 as the length of a line drawn from the anterior iris to the corneal endothelium perpendicular to a line drawn along the trabecular meshwork at a given distance from the scleral spur and (b) trabecular-iris space area at 500 μm from the scleral spur (TISA500), defined by Radhakrishnan et al8, 16 as a trapezoidal area with the following boundaries: anteriorly, AOD500; posteriorly, a line drawn from the scleral spur perpendicular to the plane of the inner scleral wall to the opposing iris; superiorly, the inner corneo-scleral wall; and inferiorly, the iris surface.

Gonioscopy

All subjects underwent gonioscopy using a Zeiss style four-mirror gonioscopy lens (Ocular Instruments Inc., Bellevue, WA, USA) by a glaucoma specialist (RJ), who was masked to the SD-ASOCT and Scheimpflug images. Gonioscopy examinations (at × 16 magnification) were performed in a dark room (1 lux using digital light meter) using a 1-mm beam reduced to a very narrow slit with minimal illumination to allow adequate visualization of the structures, while taking care to avoid light falling on the pupil or accidental indentation. Slight tilting of the gonioscopy lens was permitted in an attempt to gain a view over the convexity of the iris. The vertical beam was offset horizontally for assessing the superior and inferior angles, and was offset vertically for the nasal and temporal angles. The procedure was performed without a mydriatic or miotic agent, and a topical anesthetic was applied before examination. Shaffer grading system24, 25 was used and a narrow angle was defined as Shaffer grade ≤1 in all four quadrants.16, 26

Statistical analysis

If both eyes were eligible for the study, only the right eyes of the subjects were analyzed. Temporal and nasal angles were analyzed separately for each eye. With the exception of age, other parameters did not conform to a Guassian distribution (P<0.05; Kolmogorov–Smirnov test with a Lilliefors significance correction). As data was not normally distributed, Mann–Whitney U-test for unpaired data was used to determine differences between groups. Bonferroni correction was used to adjust P-values for multiple comparisons. The correlation between gonioscopy and ACV measured by Scheimpflug imaging was calculated using the Spearman's correlation coefficient. Sensitivity, specificity, positive likelihood ratio (LR), negative LR, positive predictive value (PPV), negative predictive value (NPV), and the receiver operating characteristic (ROC) curves for each of the SD-ASOCT and Scheimpflug imaging parameters for identifying narrow angles were calculated using gonioscopy as the reference. The area under the ROC curves (AUC) was compared using a nonparametric approach described by Delong et al.27 Optimal cutoff levels were determined using the Youden index (J);28, 29, 30 J=(sensitivity+specificity−1) for all possible cutoff values. The parameter value with the maximum Youden index was used as the cutoff value.

PPV and NPV were calculated using a prevalence of 8% from this study. We also calculated PPV and NPV using population prevalences of 2.75 and 0.88% for PAC and PACG, respectively, as determined by a recent population-based study in India31 to assess the performance of ACV. Statistical analysis was carried out using SPSS 17.0 (SPSS, Chicago, IL, USA) and MedCalc version 9.5.0.0 (MedCalc Software, Mariakerke, Belgium). Statistical significance was assumed at the P<0.05 level.

Results

A total of 300 consecutive participants underwent imaging and among them 35 subjects (11.67%) were excluded because of an undetectable scleral spur on SD-ASOCT. None of the patients were excluded because they could not undergo gonioscopy, SD-ASOCT, or Pentacam examinations. Two hundred and sixty-five eyes of 265 subjects, 136 females (51.3%) and 129 males (48.7%), were included. The mean age was 55.2±5.1 years (range 40–82 years). Mean central corneal thickness (CCT) was 539.9±37.6 μm. Twenty-eight eyes (10.6%) were classified as having narrow angles on gonioscopy (Shaffer ≤1 in all four quadrants). Table 1 describes the distribution of gonioscopic grades using the Shaffer scale. The mean BCVA was 0.10±0.20. The mean spherical equivalent of the manifest refraction was −0.53±3.09 diopters (D) and the mean IOP was 16.2±2.8 mm Hg. Table 2 describes the clinical characteristics of the study population. None of the patients had glaucoma. Of the 28 eyes with narrow angles, 4 were PAC and 10 were PAC suspects.

Table 1 Distribution of gonioscopic grade classified Shaffer scale in the study population (n=265)
Table 2 Clinical characteristics of the study population (n=265)

There was a low but significant correlation (P<0.001) between ACV calculated using Scheimpflug imaging and Gonioscopy grading for the temporal (r=0.204), superior (r=0.251), nasal (r=0.213), and inferior angles (r=0.236), and using mean gonioscopy grade (r=0.256). ACV measurements also had a low but significant correlation with SD-ASOCT-derived TISA500 for the nasal (r=0.135, P=0.029) and temporal (r=0.160, P=0.009) quadrants. (Figure 1) Similarly, there was a moderately positive correlation between ACV and AOD500 for the nasal (r=0.498, P<0.001) and temporal (r=0.517, P<0.001) quadrants (Figure 2). We did not find any correlation between ACV and age, IOP, or CCT (P>0.05 for each).

Figure 1
figure 1

Scatter plot demonstrating correlation of TISA500 (nasal and temporal) versus ACV. TISA, trabecular-iris space area at 500 μm from the scleral spur (mm2); ACV, anterior chamber volume using Scheimpflug (mm3); SD-ASOCT, spectral domain anterior segment optical coherence tomography.

Figure 2
figure 2

Scatter plot demonstrating correlation of AOD500 (nasal and temporal) versus ACV. AOD, angle opening distance at 500 μm from the scleral spur (mm); ACV, anterior chamber volume using Scheimpflug (mm3); SD-ASOCT, spectral domain anterior segment optical coherence tomography.

ROC curves for predicting narrow angles were constructed for ACV, ACD, TISA500, and AOD500 using gonioscopy as the reference (Figure 3). Comparing the sensitivity and specificity at different values of ACV, we found that using a cutoff of 113 mm3, ACV had 90% sensitivity and 88.19% specificity in detecting narrow angles with an AUC=0.935 (95% confidence interval (CI), 0.898–0.761). ACV outperformed AOD500 nasal (AUC=0.761; 95% CI, 0.705–0.811; P<0.001), AOD500 temporal (AUC=0.808; 95% CI, 0.755–0.854; P<0.001), TISA500 nasal (AUC=0.756; 95% CI, 0.7–0.807; P<0.001), and TISA500 temporal (AUC=0.738; 95%CI, 0.681–0.790; P<0.001) in detecting narrow angles. Although ACV had a higher AUC than ACD (AUC=0.880; 95% CI, 0.835–0.917), the difference was not significant (P=0.06) (Table 3). The cutoff values and the corresponding sensitivity, specificity, positive LR, negative LR, PPV and NPV for ACV, ACD, AOD500, and TISA500 (nasal and temporal) are presented in Table 3.

Figure 3
figure 3

Graph showing receiver operating characteristic curves of different parameters from Scheimpflug imaging (ACV and ACD) and SD-ASOCT (AOD500 and TISA500) of detecting narrow angles. Using a cutoff of 113 mm3, ACV had 88.29% sensitivity and 88.19% specificity in detecting narrow angles (Shaffer grade ≤1). AOD500, angle opening distance at 500 μm from the scleral spur; TISA500, trabecular-iris space area at 500 μm from the scleral spur; ACV, anterior chamber volume using Scheimpflug imaging; ACD, anterior chamber depth using Scheimpflug imaging.

Table 3 Sensitivity, specificity, positive predictive value (PPV), negative predictive value (NPV), for Scheimpflug imaging and SD-ASOCT parameters compared with gonioscopy for identifying narrow angles and area under the receiver operating characteristic curve (AUC) for parameters measured with Scheimpflug imaging and SD-ASOCT in detecting narrow angles using gonioscopy as the reference standard

Using ACV (with a cutoff of 113 mm3), the PPV for detecting narrow angles was 42.9% (95% CI, 27.7–59), while the NPV was 99% (95% CI, 96.6–99.9), using an 8% prevalence of narrow angles from this study. The PPV for detecting PAC (using a population prevalence of 2.75%31) was 19.6% (95% CI, 7.7–37.8) and the NPV was 99.7% (95% CI, 97.7–99.9). For detecting PACG (using a population prevalence of 0.88%31), the PPV was 7.1% (95% CI, 1.0–23.9) and the NPV was 99.9% (95% CI, 98.1–99.7) At the cutoff of 113 mm3, ACV had a positive LR of 8.63 (95% CI, 7.4–10.0) and a negative LR of 0.11 (95% CI, 0.03–0.4) for detecting narrow angles.

We also used another definition for narrow angles as grade ≤2 in all quadrants. Using this definition, ACV had an AUC of 0.901 and an 89.1% sensitivity and 87.2% specificity in detecting narrow angles. (95% CI, 0.822–0.931). Again ACV (AUC=0.901) outperformed AOD500 nasal (AUC=0.761; 95% CI, 0.685–0.761; P<0.001), AOD500 temporal (AUC=0.788; 95% CI, 0.722–0.814; P<0.001), TISA500 nasal (AUC=0.726; 95% CI, 0.664–0.798; P<0.001), and TISA500 temporal (AUC=0.722; 95%CI, 0.676–0.787; P<0.001) in detecting narrow angles, but was similar to ACD (AUC=0.864; 95% CI, 0.815–0.907).

Discussion

The ideal community-based screening test should not only be clinician-independent, quick and non-invasive, with a high specificity and sensitivity but also be practical and of an affordable cost. Several authors have compared ASOCT with gonioscopy as the reference and have shown sensitivities of upto 98% in detecting narrow angles.32, 33

Angle classification in ASOCT hinges on accurate localization of the scleral spur,34 as it is used as the reference point for the relative position of trabecular meshwork. Localization of the scleral spur can be difficult, with closed angles and in the superior and inferior angles.7, 16 Even with identical criteria, intraobserver variance in identifying the spur location can be high. In our study, the rate of undetectable sclera spur was 11.67%.

Fully automated edge detection algorithms for the ACA are being developed in an attempt to address this issue and have been described in ultrasound biomicroscopy (UBM) images,35 but are not commercially available on SD-ASOCT. ACV measurements using Scheimpflug imaging are automated and do not require manual identification of the scleral spur. Kurita et al20 recently reported that ACV had the highest discriminating ability (AUC=0.943) in screening for PAC and PAC suspects compared with UBM parameters. Using a larger sample size, we compared the diagnostic performance of ACV with SD-ASOCT angle measurements, and found that ACV had the highest discriminating ability (AUC=0.935) outperforming AOD500 and TISA500 in detecting narrow angles. We hypothesize that one of the reasons why ACV outperformed SD-ASOCT parameters is because of the ability of the rotating Scheimpflug camera to capture 50 slices through the anterior segment, and using this data to calculate ACV. Angle parameter calculations using SD-ASOCT are calculated, and are limited to single cross-sectional images of the nasal and temporal quadrants. Although in most cases this may represent the overall nature of the angle as seen gonioscopically, there are sectoral variations in eyes with uneven angle configuration,17 which will be missed on single SD-ASOCT images.

Our study has several limitations. The study group was not population based and had a relatively small sample size of 265 eyes, limited to Indian eyes. Validating these findings in a more diverse population-based study would be useful. For screening purposes, ideally the specificity should be in the high 90s,36 a value which none of the parameters in this study achieved. Scheimpflug photography has its limitations in that it cannot fully visualize the entire angle.19 This leads to the location of the angle recess being estimated, and extrapolation of data by the Pentacam software to calculate the ACV.37 Additionally, as it uses visible light to image the angle, it may lead to altered angle configuration.38 Although repeatability of ACV measurements has been demonstrated with an intraclass correlation coefficient of 0.991,39 there are concerns that as ACV is reported with no decimal places, it may be affected by rounding error.40 Our results are not applicable to the superior and inferior angle quadrants, as these were not imaged using SD-ASOCT. Also, we measured the parameters only under dark conditions and cannot report on the role of dynamic changes induced by different lighting conditions. Use of external fixation to center the iridocorneal angle in the instrument's field of view limits the precision for absolute standardization and centration. The shorter wavelength used in SD-ASOCT compared with TD-ASOCT does not adequately penetrate highly scattering tissue such as the sclera or iris, and thus may not be optimal for ACA imaging. Although the increased resolution of SD-ASOCT allows for identification of the Schwalbe's line, Schlemm's canal, and trabecular meshwork in some eyes, the depth and width of view provided by SD-ASOCT precludes imaging of the entire angle up to the iris root,41, 19 which is most important when screening for narrow angles. Angle images acquired using the SD-ASOCT used in this study might be prone to distortion, as they are not dewarped.

Although correlations between ACV, gonioscopy grading and SD-ASOCT were statistically significant, most of them represent low (r<0.30) to moderate associations (r<0.50).42, 43 The dynamic nature of the anterior chamber might be partially responsible for the low correlations as also the fact that SD-ASOCT images were captured in just one cross-section through the angle in each quadrant, which may not be representative of the entire angle configuration in that quadrant.

Evidence-based medicine guidelines have suggested the appropriateness of LR to judge the clinical utility of test results.44, 45 The large positive LR (8.63; 95% CI, 7.4–10) suggests that an eye detected as having narrow angles using ACV (≤113 mm3) will have a moderate chance of having narrow angles on gonioscopy. The negative LR (0.11; 95% CI, 0.03–0.40) suggests that a negative result on ACV (>113 mm3) could also be used with moderate confidence to rule out narrow angles. However, calculations of LRs depend on gonioscopy as the reference, and its limitations as an imperfect standard are well known. Even experienced examiners have reported to have only a moderate agreement in determining the angle width.46, 47

In conclusion, ACV measurements using the commercially available, automated software demonstrated a high discriminative value in screening for narrow angles. Improvement in image analysis algorithms could allow for the development of ‘cutoff’ points for discriminating narrow angles, or at least specific thresholds used to trigger a referral for examination. Only longitudinal prospective studies can accurately evaluate the predictive ability of these parameters by determining whether eyes classified as narrow angles, only by Scheimpflug or SD-ASOCT imaging, are indeed at risk of developing PACG.34 While keeping these limitations in mind, this study provides data demonstrating the potential for using ACV as a screening tool for narrow angles in an Asian Indian population.