Abstract
Background: The aim of this comparative study was to gain subjective and objective data to determine for which operative tasks three-dimensional (3-D) vision systems are superior to two-dimensional (2-D) systems and to demonstrate any advantages or disadvantages of 3-D systems.
Methods: A model with five standardized tasks including sewing and knotting was developed to objectively measure performance times and to count technical faults. In our training center for minimally invasive surgery, surgeons involved in basic and advanced laparoscopic courses trained using both 2-D and 3-D vision systems. They subsequently completed analog scale questionnaires to record a subjective impression of comparative ease of operation tasks under 2-D and 3-D vision and to identify perceived deficiencies in the 3-D system.
Results: Compared to 2-D vision, the objective performance time was significantly shorter and significantly less mistakes were made using 3-D vision. All operative tasks were subjectively judged significantly easier under 3-D vision.
Conclusions: Users with a normal capability for spatial perception can perform standard tasks more quickly and safely using 3-D vision, and a greater benefit is apparent for more complicated surgical maneuvers.
Similar content being viewed by others
Author information
Authors and Affiliations
Additional information
Received: 29 January 1997/Accepted: 18 September 1997
Rights and permissions
About this article
Cite this article
van Bergen, P., Kunert, W., Bessell, J. et al. Comparative study of two-dimensional and three-dimensional vision systems for minimally invasive surgery. Surg Endosc 12, 948–954 (1998). https://doi.org/10.1007/s004649900754
Published:
Issue Date:
DOI: https://doi.org/10.1007/s004649900754