Clinical evaluation of complete solo surgery with the “ViKY®” robotic laparoscope manipulator

  • New Technology
  • Published:
Surgical Endoscopy Aims and scope Submit manuscript

Abstract

Background

Advancement in both surgical technique and medical equipment has enabled solo surgery. ViKY® Endoscope Positioning System (ViKY®) is a robotic system that remotely controls an endoscope and provides direct vision control to the surgeon. Here, we report our experience with ViKY®-assisted solo surgery.

Methods

We retrospectively examined 25 cases of solo surgery TAPP with ViKY®. ViKY® was setup by the surgeon alone, and the setup duration was determined as the time at which the side rail was positioned and that when the endoscope was installed. For assessing the control unit, the number of false movements was counted. We compared the operative results between ViKY®-assisted solo surgery TAPP and the conventional method with an assistant.

Results

The average time to set up ViKY® was 7.9 min. The average number of commands for ViKY® during surgery was 98.3, and the average number of errors and no response of control unit was 7.9. The mean duration of surgery was 136 min for the ViKY® group, including the setup time, and 117 min for the conventional method. No case required an assistant during the operation. There was also no difference between the two groups with regard to postoperative complications and the rate of recurrence.

Conclusions

ViKY® proved reliable in recognizing orders with very few failures, and the operations were performed safely and were comparable to the conventional operations with assistants. Solo surgery with ViKY® was beneficial in this clinical evaluation.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Institutional subscriptions

Fig. 1
Fig. 2

Similar content being viewed by others

References

  1. Mizota T, Watanabe Y, Madani A, Takemoto N, Yamada H, Poudel S, Miyasaka Y, Kurashima Y (2015) Liquid-injection for preperitoneal dissection of transabdominal preperitoneal (TAPP) inguinal hernia repair. Surg Endosc 29:516–520

    Article  PubMed  Google Scholar 

  2. Anthony R (2004) Robotic surgery: a current perspective. Ann Surg 239:14–21

    Article  Google Scholar 

  3. Sackier JM, Wang Y (1994) Robotically assisted laparoscopic surgery. Surg Endosc 8:63–66

    Article  CAS  PubMed  Google Scholar 

  4. Schurr MO, Arezzo A, Neisius B, Rininsland H, Hilzinger HU, Dorn J, Roth K, Buess GF (1999) Trocar and instrument positioning system TISKA. An assist device for endoscopic solo surgery. Surg Endosc 13:528–531

    Article  CAS  PubMed  Google Scholar 

  5. Aiono S, Gilbert JM, Soin B, Finlay PA, Gordan A (2002) Controlled trial of the introduction of a robotic camera assistant (EndoAssist) for laparoscopic cholecystectomy. Surg Endosc 16:1267–1270

    Article  CAS  PubMed  Google Scholar 

  6. Polet R, Donnez J (2008) Using a laparoscope manipulator (LAPMAN) in laparoscopic gynecological surgery. Surg Technol Int 17:187–191

    PubMed  Google Scholar 

  7. Yamada K, Kato S (2008) Robot-assisted thoracoscopic lung resection aimed at solo surgery for primary lung cancer. Gen Thorac Cardiovasc Surg 56:292–294

    Article  PubMed  Google Scholar 

  8. Gillen S, Pletzer B, Heiligensetzer A, Wolf P, Kleeff J, Feussner H, Fürst A (2014) Solo-surgical laparoscopic cholecystectomy with a joystick-guided camera device: a case–control study. Surg Endosc 28:164–170

    Article  PubMed  Google Scholar 

  9. John H (2011) Dirckx, Robotic surgery, The SUM Program Obstetrics/Gynecology Advanced Medical Transcription Unit, 2nd edn. Health Profession Institute, http://www.hpisum.com/Dirckx,%20Robotic%20Surgery.pdf. Accessed 9 June 2016

  10. Barbash GI, Glied SA (2010) New technology and health care costs—the case of robot-assisted surgery. N Engl J Med 363(8):701–704

    Article  CAS  PubMed  Google Scholar 

  11. Alemzadeh H, Iyer RK, Kalbarczyk Z, Leveson N, Raman J (2016) Adverse, events in robotic surgery: a retrospective study of 14 years of FDA data. http://arxiv.org/abs/1507.03518. Accessed 9 June 2016

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Masahiro Takahashi.

Ethics declarations

Disclosures

Masahiro Takahashi, Masanori Takahashi, Naoto Nishinari, Hideki Matsuya, Tsutomu Tosha, Yukihiro Minagawa, Osamu Shimooki, and Tadashi Abe have no conflicts of interest or financial ties to disclose.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Takahashi, M., Takahashi, M., Nishinari, N. et al. Clinical evaluation of complete solo surgery with the “ViKY®” robotic laparoscope manipulator. Surg Endosc 31, 981–986 (2017). https://doi.org/10.1007/s00464-016-5058-8

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00464-016-5058-8

Keywords

Navigation