Effects of AV-delay optimization on hemodynamic parameters in patients with VDD pacemakers

verfasst von: Konstantin A. Krychtiuk, MD, Michael Nürnberg, MD, Romana Volker, MD, Linda Pachinger, MD, Rudolf Jarai, MD, Matthias K. Freynhofer, MD, Johann Wojta, PhD, Kurt Huber, MD, Thomas W. Weiss, MD, PhD

Erschienen in: Wiener klinische Wochenschrift | Ausgabe 9-10/2014

Einloggen, um Zugang zu erhalten

Summary

Aim

Atrioventricular (AV) delay optimization improves hemodynamics and clinical parameters in patients treated with cardiac resynchronization therapy and dual-chamber-pacemakers (PM). However, data on optimizing AV delay in patients treated with VDD-PMs are scarce. We, therefore, investigated the acute and chronic effects of AV delay optimization on hemodynamics in patients treated with VDD-PMs due to AV-conduction disturbances.

Methods

In this prospective, single-center interventional trial, we included 64 patients (38 men, 26 women, median age: 77 (70–82) years) with implanted VDD-PM. AV-delay optimization was performed using a formula based on the surface electrocardiogram (ECG). Hemodynamic parameters (stroke volume (SV), cardiac output (CO), heart rate (HR), and blood pressure (BP)) were measured at baseline and follow-up after 3 months using impedance cardiography.

Results

Using an ECG formula for AV-delay optimization, the AV interval was decreased from 180 (180–180) to 75 (75–100) ms. At baseline, AV-delay optimization led to a significant increase of both SV (71.3 ± 15.8 vs. 55.3 ± 12.7 ml, p < 0.001, for optimized AV delay vs. nominal AV interval, respectively) and CO (5.1 ± 1.4 vs. 3.9 ± 1.0 l/min, p < 0.001), while HR and BP remained unchanged. At follow-up, the improvement in CO remained stable (4.9 ± 1.3 l/min, p = 0.09), while SV slightly, but significantly, decreased (to 65.1 ± 17.6, p < 0.01).

Conclusion

AV-delay optimization in patients treated with VDD-PMs exhibits immediate beneficial effects on hemodynamic parameters that are sustained for 3 months.
Literatur
1.
Zurück zum Zitat Vardas PE, Auricchio A, Blanc JJ, Daubert JC, Drexler H, Ector H, et al. Guidelines for cardiac pacing and cardiac resynchronization therapy. The task force for cardiac pacing and cardiac resynchronization therapy of the European Society of Cardiology. Developed in collaboration with the European Heart Rhythm Association. Europace. 2007;9:959–98. Vardas PE, Auricchio A, Blanc JJ, Daubert JC, Drexler H, Ector H, et al. Guidelines for cardiac pacing and cardiac resynchronization therapy. The task force for cardiac pacing and cardiac resynchronization therapy of the European Society of Cardiology. Developed in collaboration with the European Heart Rhythm Association. Europace. 2007;9:959–98.
2.
Zurück zum Zitat Marchandise S, Scavee C, le Polain de Waroux JB, de Meester C, Vanoverschelde JL, Debbas N. Long-term follow-up of DDD and VDD pacing: a prospective non-randomized single-centre comparison of patients with symptomatic atrioventricular block. Europace. 2012;14:496–501. Marchandise S, Scavee C, le Polain de Waroux JB, de Meester C, Vanoverschelde JL, Debbas N. Long-term follow-up of DDD and VDD pacing: a prospective non-randomized single-centre comparison of patients with symptomatic atrioventricular block. Europace. 2012;14:496–501.
3.
Zurück zum Zitat Schaer BA, Weinbacher M, Zellweger MJ, Sticherling C, Osswald S. Value of VDD-pacing systems in patients with atrioventricular block: experience over a decade. Int J Cardiol. 2007;122:239–43. Schaer BA, Weinbacher M, Zellweger MJ, Sticherling C, Osswald S. Value of VDD-pacing systems in patients with atrioventricular block: experience over a decade. Int J Cardiol. 2007;122:239–43.
4.
Zurück zum Zitat Huang M, Krahn AD, Yee R, Klein GJ, Skanes AC. Optimal pacing for symptomatic AV block: a comparison of VDD and DDD pacing. Pacing Clin Electrophysiol. 2004;27:19–23. Huang M, Krahn AD, Yee R, Klein GJ, Skanes AC. Optimal pacing for symptomatic AV block: a comparison of VDD and DDD pacing. Pacing Clin Electrophysiol. 2004;27:19–23.
5.
Zurück zum Zitat Wiegand UK, Potratz J, Bode F, Schreiber R, Bonnemeier H, Peters W, et al. Cost-effectiveness of dual-chamber pacemaker therapy: does single lead VDD pacing reduce treatment costs of atrioventricular block? Eur Heart J. 2001;22:174–80. Wiegand UK, Potratz J, Bode F, Schreiber R, Bonnemeier H, Peters W, et al. Cost-effectiveness of dual-chamber pacemaker therapy: does single lead VDD pacing reduce treatment costs of atrioventricular block? Eur Heart J. 2001;22:174–80.
6.
Zurück zum Zitat Braunwald E, Frahm CJ. Studies on Starling’s law of the heart. IV. Observations on the hemodynamic functions of the left atrium in man. Circulation. 1961;24:633–42. Braunwald E, Frahm CJ. Studies on Starling’s law of the heart. IV. Observations on the hemodynamic functions of the left atrium in man. Circulation. 1961;24:633–42.
7.
Zurück zum Zitat Auricchio A, Stellbrink C, Block M, Sack S, Vogt J, Bakker P, et al. Effect of pacing chamber and atrioventricular delay on acute systolic function of paced patients with congestive heart failure. The pacing therapies for Congestive Heart Failure Study Group. The Guidant Congestive Heart Failure Research Group. Circulation. 1999;99:2993–3001. Auricchio A, Stellbrink C, Block M, Sack S, Vogt J, Bakker P, et al. Effect of pacing chamber and atrioventricular delay on acute systolic function of paced patients with congestive heart failure. The pacing therapies for Congestive Heart Failure Study Group. The Guidant Congestive Heart Failure Research Group. Circulation. 1999;99:2993–3001.
8.
Zurück zum Zitat Auricchio A, Stellbrink C, Sack S, Block M, Vogt J, Bakker P, et al. Long-term clinical effect of hemodynamically optimized cardiac resynchronization therapy in patients with heart failure and ventricular conduction delay. J Am Coll Cardiol. 2002;39:2026–33. Auricchio A, Stellbrink C, Sack S, Block M, Vogt J, Bakker P, et al. Long-term clinical effect of hemodynamically optimized cardiac resynchronization therapy in patients with heart failure and ventricular conduction delay. J Am Coll Cardiol. 2002;39:2026–33.
9.
Zurück zum Zitat Hardt SE, Yazdi SH, Bauer A, Filusch A, Korosoglou G, Hansen A, et al. Immediate and chronic effects of AV-delay optimization in patients with cardiac resynchronization therapy. Int J Cardiol. 2007;115:318–25. Hardt SE, Yazdi SH, Bauer A, Filusch A, Korosoglou G, Hansen A, et al. Immediate and chronic effects of AV-delay optimization in patients with cardiac resynchronization therapy. Int J Cardiol. 2007;115:318–25.
10.
Zurück zum Zitat Adlbrecht C, Hulsmann M, Gwechenberger M, Graf S, Wiesbauer F, Strunk G, et al. Electrical optimization of cardiac resynchronization in chronic heart failure is associated with improved clinical long-term outcome. Eur J Clin Invest. 2010;40:678–84. Adlbrecht C, Hulsmann M, Gwechenberger M, Graf S, Wiesbauer F, Strunk G, et al. Electrical optimization of cardiac resynchronization in chronic heart failure is associated with improved clinical long-term outcome. Eur J Clin Invest. 2010;40:678–84.
11.
Zurück zum Zitat Kindermann M, Frohlig G, Doerr T, Schieffer H. Optimizing the AV delay in DDD pacemaker patients with high degree AV block: mitral valve Doppler versus impedance cardiography. Pacing Clin Electrophysiol. 1997;20:2453–62. Kindermann M, Frohlig G, Doerr T, Schieffer H. Optimizing the AV delay in DDD pacemaker patients with high degree AV block: mitral valve Doppler versus impedance cardiography. Pacing Clin Electrophysiol. 1997;20:2453–62.
12.
Zurück zum Zitat Toda N, Ishikawa T, Nozawa N, Kobayashi I, Ochiai H, Miyamoto K, et al. Doppler index and plasma level of atrial natriuretic hormone are improved by optimizing atrioventricular delay in atrioventricular block patients with implanted DDD pacemakers. Pacing Clin Electrophysiol. 2001;24:1660–3. Toda N, Ishikawa T, Nozawa N, Kobayashi I, Ochiai H, Miyamoto K, et al. Doppler index and plasma level of atrial natriuretic hormone are improved by optimizing atrioventricular delay in atrioventricular block patients with implanted DDD pacemakers. Pacing Clin Electrophysiol. 2001;24:1660–3.
13.
Zurück zum Zitat Crystal E, Ovsyshcher IE. Cardiac output-based versus empirically programmed AV interval–how different are they? Europace. 1999;1:121–5. Crystal E, Ovsyshcher IE. Cardiac output-based versus empirically programmed AV interval–how different are they? Europace. 1999;1:121–5.
14.
Zurück zum Zitat Liebold A, Rodig G, Merk J, Birnbaum DE. Short atrioventricular delay dual-chamber pacing early after coronary artery bypass grafting in patients with poor left ventricular function. J Cardiothorac Vasc Anesth. 1998;12:284–7. Liebold A, Rodig G, Merk J, Birnbaum DE. Short atrioventricular delay dual-chamber pacing early after coronary artery bypass grafting in patients with poor left ventricular function. J Cardiothorac Vasc Anesth. 1998;12:284–7.
15.
Zurück zum Zitat Leman RB, Kratz JM. Radionuclide evaluation of dual chamber pacing: comparison between variable AV intervals and ventricular pacing. Pacing Clin Electrophysiol. 1985;8:408–14. Leman RB, Kratz JM. Radionuclide evaluation of dual chamber pacing: comparison between variable AV intervals and ventricular pacing. Pacing Clin Electrophysiol. 1985;8:408–14.
16.
Zurück zum Zitat Whinnett ZI, Davies JE, Willson K, Chow AW, Foale RA, Davies DW, et al. Determination of optimal atrioventricular delay for cardiac resynchronization therapy using acute non-invasive blood pressure. Europace. 2006;8:358–66. Whinnett ZI, Davies JE, Willson K, Chow AW, Foale RA, Davies DW, et al. Determination of optimal atrioventricular delay for cardiac resynchronization therapy using acute non-invasive blood pressure. Europace. 2006;8:358–66.
17.
Zurück zum Zitat Zuber M, Toggweiler S, Roos M, Kobza R, Jamshidi P, Erne P. Comparison of different approaches for optimization of atrioventricular and interventricular delay in biventricular pacing. Europace. 2008;10:367–73. Zuber M, Toggweiler S, Roos M, Kobza R, Jamshidi P, Erne P. Comparison of different approaches for optimization of atrioventricular and interventricular delay in biventricular pacing. Europace. 2008;10:367–73.
18.
Zurück zum Zitat Ritter P, Dib J, Mahaux V. New method for determining the optimal atrio-ventricular delay in DDD mode for complete atrio-ventricular block. Pacing Clin Electrophysiol. 2008;18:237 (abstract). Ritter P, Dib J, Mahaux V. New method for determining the optimal atrio-ventricular delay in DDD mode for complete atrio-ventricular block. Pacing Clin Electrophysiol. 2008;18:237 (abstract).
19.
Zurück zum Zitat Meluzin J, Novak M, Mullerova J, Krejci J, Hude P, Eisenberger M, et al. A fast and simple echocardiographic method of determination of the optimal atrioventricular delay in patients after biventricular stimulation. Pacing Clin Electrophysiol. 2004;27:58–64. Meluzin J, Novak M, Mullerova J, Krejci J, Hude P, Eisenberger M, et al. A fast and simple echocardiographic method of determination of the optimal atrioventricular delay in patients after biventricular stimulation. Pacing Clin Electrophysiol. 2004;27:58–64.
20.
Zurück zum Zitat Ovsyshcher I, Gross JN, Blumberg S, Furman S. Precision of impedance cardiography measurements of cardiac output in pacemaker patients. Pacing Clin Electrophysiol. 1992;15:1923–6. Ovsyshcher I, Gross JN, Blumberg S, Furman S. Precision of impedance cardiography measurements of cardiac output in pacemaker patients. Pacing Clin Electrophysiol. 1992;15:1923–6.
21.
Zurück zum Zitat Ovsyshcher I, Furman S. Impedance cardiography for cardiac output estimation in pacemaker patients: review of the literature. Pacing Clin Electrophysiol. 1993;16:1412–22. Ovsyshcher I, Furman S. Impedance cardiography for cardiac output estimation in pacemaker patients: review of the literature. Pacing Clin Electrophysiol. 1993;16:1412–22.
22.
Zurück zum Zitat Heinroth KM, Elster M, Nuding S, Schlegel F, Christoph A, Carter J, et al. Impedance cardiography: a useful and reliable tool in optimization of cardiac resynchronization devices. Europace. 2007;9:744–50. Heinroth KM, Elster M, Nuding S, Schlegel F, Christoph A, Carter J, et al. Impedance cardiography: a useful and reliable tool in optimization of cardiac resynchronization devices. Europace. 2007;9:744–50.
23.
Zurück zum Zitat Drazner MH, Thompson B, Rosenberg PB, Kaiser PA, Boehrer JD, Baldwin BJ, et al. Comparison of impedance cardiography with invasive hemodynamic measurements in patients with heart failure secondary to ischemic or nonischemic cardiomyopathy. Am J Cardiol. 2002;89:993–5. Drazner MH, Thompson B, Rosenberg PB, Kaiser PA, Boehrer JD, Baldwin BJ, et al. Comparison of impedance cardiography with invasive hemodynamic measurements in patients with heart failure secondary to ischemic or nonischemic cardiomyopathy. Am J Cardiol. 2002;89:993–5.
24.
Zurück zum Zitat Sageman WS, Riffenburgh RH, Spiess BD. Equivalence of bioimpedance and thermodilution in measuring cardiac index after cardiac surgery. J Cardiothorac Vasc Anesth. 2002;16:8–14. Sageman WS, Riffenburgh RH, Spiess BD. Equivalence of bioimpedance and thermodilution in measuring cardiac index after cardiac surgery. J Cardiothorac Vasc Anesth. 2002;16:8–14.
25.
Zurück zum Zitat Koglek W, Kowalski KW. A simple method for AV-delay determination in dual chamber pacemakers (german). Herzschr Elekrophys. 2000;11:244–53. Koglek W, Kowalski KW. A simple method for AV-delay determination in dual chamber pacemakers (german). Herzschr Elekrophys. 2000;11:244–53.
26.
Zurück zum Zitat Strohmer B, Pichler M, Froemmel M, Migschitz M, Hintringer F. Evaluation of atrial conduction time at various sites of right atrial pacing and influence on atrioventricular delay optimization by surface electrocardiography. Pacing Clin Electrophysiol. 2004;27:468–74. Strohmer B, Pichler M, Froemmel M, Migschitz M, Hintringer F. Evaluation of atrial conduction time at various sites of right atrial pacing and influence on atrioventricular delay optimization by surface electrocardiography. Pacing Clin Electrophysiol. 2004;27:468–74.
27.
Zurück zum Zitat Braun MU, Schnabel A, Rauwolf T, Schulze M, Strasser RH. Impedance cardiography as a noninvasive technique for atrioventricular interval optimization in cardiac resynchronization therapy. J Interv Card Electrophysiol. 2005;13:223–9. Braun MU, Schnabel A, Rauwolf T, Schulze M, Strasser RH. Impedance cardiography as a noninvasive technique for atrioventricular interval optimization in cardiac resynchronization therapy. J Interv Card Electrophysiol. 2005;13:223–9.
28.
Zurück zum Zitat Bogaard MD, Meine M, Tuinenburg AE, Maskara B, Loh P, Doevendans PA. Cardiac resynchronization therapy beyond nominal settings: who needs individual programming of the atrioventricular and interventricular delay? Europace. 2012;14:1746–53. Bogaard MD, Meine M, Tuinenburg AE, Maskara B, Loh P, Doevendans PA. Cardiac resynchronization therapy beyond nominal settings: who needs individual programming of the atrioventricular and interventricular delay? Europace. 2012;14:1746–53.
29.
Zurück zum Zitat Manisty CH, Al-Hussaini A, Unsworth B, Baruah R, Pabari PA, Mayet J, et al. The acute effects of changes to AV delay on BP and stroke volume: potential implications for design of pacemaker optimization protocols. Circ Arrhythm Electrophysiol. 2012;5:122–30. Manisty CH, Al-Hussaini A, Unsworth B, Baruah R, Pabari PA, Mayet J, et al. The acute effects of changes to AV delay on BP and stroke volume: potential implications for design of pacemaker optimization protocols. Circ Arrhythm Electrophysiol. 2012;5:122–30.
30.
Zurück zum Zitat Sawhney NS, Waggoner AD, Garhwal S, Chawla MK, Osborn J, Faddis MN. Randomized prospective trial of atrioventricular delay programming for cardiac resynchronization therapy. Heart Rhythm. 2004;1:562–7. Sawhney NS, Waggoner AD, Garhwal S, Chawla MK, Osborn J, Faddis MN. Randomized prospective trial of atrioventricular delay programming for cardiac resynchronization therapy. Heart Rhythm. 2004;1:562–7.
31.
Zurück zum Zitat Morales MA, Startari U, Panchetti L, Rossi A, Piacenti M. Atrioventricular delay optimization by doppler-derived left ventricular dP/dt improves 6-month outcome of resynchronized patients. Pacing Clin Electrophysiol. 2006;29:564–8. Morales MA, Startari U, Panchetti L, Rossi A, Piacenti M. Atrioventricular delay optimization by doppler-derived left ventricular dP/dt improves 6-month outcome of resynchronized patients. Pacing Clin Electrophysiol. 2006;29:564–8.
32.
Zurück zum Zitat Ritter PH, Fischer W. Cardiac Pacing in Clinical Practice. Springer Verlag; 1998. Ritter PH, Fischer W. Cardiac Pacing in Clinical Practice. Springer Verlag; 1998.
Metadaten
Titel
Effects of AV-delay optimization on hemodynamic parameters in patients with VDD pacemakers
verfasst von
Konstantin A. Krychtiuk, MD
Michael Nürnberg, MD
Romana Volker, MD
Linda Pachinger, MD
Rudolf Jarai, MD
Matthias K. Freynhofer, MD
Johann Wojta, PhD
Kurt Huber, MD
Thomas W. Weiss, MD, PhD
Publikationsdatum
01.05.2014
Verlag
Springer Vienna
Erschienen in
Wiener klinische Wochenschrift / Ausgabe 9-10/2014
Print ISSN: 0043-5325
Elektronische ISSN: 1613-7671
DOI
https://doi.org/10.1007/s00508-014-0520-x

Weitere Artikel der Ausgabe 9-10/2014

Wiener klinische Wochenschrift 9-10/2014 Zur Ausgabe

images in clinical medicine

A rare tumor: tracheal Ewing’s sarcoma