For how many times can a partially absorbable monofilament mesh be sterilized? In vitro experimental study

verfasst von: O. V. Ozkan, MD, A. Aydogan, MD, S. Akkucuk, MD, M. Aydin, MD, E. Nacar, B. Ozer, S. Mistikoglu, I. Yetim, MD, A. Gokce, MD

Erschienen in: European Surgery | Ausgabe 2/2013

Einloggen, um Zugang zu erhalten

Summary

Background

We investigated the effects of hydrogen peroxide gas plasma resterilization (HPS) on partially absorbable monofilament meshes in terms of their mechanical properties and risk of infection.

Methods

A mesh was divided into small pieces which were categorized as HPS-1, HPS-2 and HPS-3 according to the number of HPS. To assess the strength of specimens, maximum load, elongation at maximum load and quantity of energy required for complete failure of the specimens were measured. Microbiological and ultrastructural analyses were also performed.

Results

There was not a significant relationship between control and HPS-1, HPS-2 groups in terms of maximum load and quantity of energy required for complete failure of the specimens. However, those parameters were statistically different between HPS-3 and control groups. We observed minor morphological changes in the HPS-3 group when compared to those of the control group. No risk of infection was detected by microbiological tests.

Conclusion

It is advisable to apply HPS to partially absorbable monofilament meshes no more than twice since sterilization for three times leads to degeneration in mesh structure and strength.
Literatur
1.
Zurück zum Zitat Collaboration EH. Mesh compared with non-mesh methods of open groin hernia repair: systematic review of randomized controlled trials. Br J Surg. 2000;87:854–9.CrossRef Collaboration EH. Mesh compared with non-mesh methods of open groin hernia repair: systematic review of randomized controlled trials. Br J Surg. 2000;87:854–9.CrossRef
2.
3.
Zurück zum Zitat Muller MAR, Osbahr DC, Murrell GA. Meshes in rotator cuff repair: a biomechanical study (TECHNIQUE). Tech shoulder elbow surg. 2003;4:50–4.CrossRef Muller MAR, Osbahr DC, Murrell GA. Meshes in rotator cuff repair: a biomechanical study (TECHNIQUE). Tech shoulder elbow surg. 2003;4:50–4.CrossRef
4.
Zurück zum Zitat Bellon JM, Rodriguez M, Garcia-Honduvilla N, Pascual G, Bujan J. Partially absorbable meshes for hernia repair offer advantages over nonabsorbable meshes. Am J Surg. 2007;194:68–74.CrossRefPubMed Bellon JM, Rodriguez M, Garcia-Honduvilla N, Pascual G, Bujan J. Partially absorbable meshes for hernia repair offer advantages over nonabsorbable meshes. Am J Surg. 2007;194:68–74.CrossRefPubMed
5.
Zurück zum Zitat Langer C, Neufang T, Kley C, Liersch T, Becker H. Central mesh recurrence after incisional hernia repair with Marlex-are the meshes strong enough? Hernia. 2001;5:164–7.CrossRefPubMed Langer C, Neufang T, Kley C, Liersch T, Becker H. Central mesh recurrence after incisional hernia repair with Marlex-are the meshes strong enough? Hernia. 2001;5:164–7.CrossRefPubMed
6.
Zurück zum Zitat Coda A, Bendavid R, Botto-Micca F, Bossotti M, Bona A. Structural alterations of prosthetic meshes in humans. Hernia. 2003;7:29–34.PubMed Coda A, Bendavid R, Botto-Micca F, Bossotti M, Bona A. Structural alterations of prosthetic meshes in humans. Hernia. 2003;7:29–34.PubMed
7.
Zurück zum Zitat Serbetci K, Kulacoglu H, Devay AO, Hasirci N. Effects of resterilization on mechanical properties of polypropylene meshes. Am J Surg. 2007;194:375–9.CrossRefPubMed Serbetci K, Kulacoglu H, Devay AO, Hasirci N. Effects of resterilization on mechanical properties of polypropylene meshes. Am J Surg. 2007;194:375–9.CrossRefPubMed
8.
Zurück zum Zitat Nagaraja PA, Shetty D. Effect of re-sterilization of surgical sutures by ethylene oxide. ANZ J Surg. 2007;77:80–3.CrossRefPubMed Nagaraja PA, Shetty D. Effect of re-sterilization of surgical sutures by ethylene oxide. ANZ J Surg. 2007;77:80–3.CrossRefPubMed
9.
Zurück zum Zitat Cobb WS, Kercher KW. Heniford BT The argument for lightweight polypropylene mesh in hernia repair. Surg Innov. 2005;12:63–9.CrossRefPubMed Cobb WS, Kercher KW. Heniford BT The argument for lightweight polypropylene mesh in hernia repair. Surg Innov. 2005;12:63–9.CrossRefPubMed
10.
Zurück zum Zitat Bellon JM, Bujan J, Contreras L, Hernando A. Integration of biomaterials implanted into abdominal wall: process of scar formation and macrophage response. Biomaterials. 1995;16:381–7.CrossRefPubMed Bellon JM, Bujan J, Contreras L, Hernando A. Integration of biomaterials implanted into abdominal wall: process of scar formation and macrophage response. Biomaterials. 1995;16:381–7.CrossRefPubMed
11.
Zurück zum Zitat Akolekar D, Kumar S, Khan LR, de Beaux AC, Nixon SJ. Comparison of recurrence with lightweight composite polypropylene mesh and heavyweight mesh in laparoscopic totally extraperitoneal inguinal hernia repair: an audit of 1,232 repairs. Hernia. 2008;12:39–43.CrossRefPubMed Akolekar D, Kumar S, Khan LR, de Beaux AC, Nixon SJ. Comparison of recurrence with lightweight composite polypropylene mesh and heavyweight mesh in laparoscopic totally extraperitoneal inguinal hernia repair: an audit of 1,232 repairs. Hernia. 2008;12:39–43.CrossRefPubMed
12.
Zurück zum Zitat Agarwal BB, Agarwal KA, Mahajan KC. Prospective double-blind randomized controlled study comparing heavy- and lightweight polypropylene mesh in totally extraperitoneal repair of inguinal hernia: early results. Surg Endosc. 2009;23:242–7.CrossRefPubMed Agarwal BB, Agarwal KA, Mahajan KC. Prospective double-blind randomized controlled study comparing heavy- and lightweight polypropylene mesh in totally extraperitoneal repair of inguinal hernia: early results. Surg Endosc. 2009;23:242–7.CrossRefPubMed
13.
Zurück zum Zitat Junge K, Rosch R, Krones CJ, Klinge U, Mertens PR, Lynen P, et al. Influence of polyglecaprone 25 (Monocryl) supplementation on the biocompatibility of a polypropylene mesh for hernia repair. Hernia. 2005;9:212–7.CrossRefPubMed Junge K, Rosch R, Krones CJ, Klinge U, Mertens PR, Lynen P, et al. Influence of polyglecaprone 25 (Monocryl) supplementation on the biocompatibility of a polypropylene mesh for hernia repair. Hernia. 2005;9:212–7.CrossRefPubMed
14.
Zurück zum Zitat Holzheimer RG. First results of Lichtenstein hernia repair with ultrapro-mesh as cost saving procedure-quality control combined with a modified quality of life questionnaire (SF- 36) in a series of ambulatory operated patients. Eur J Med Res. 2004;9:323–7.PubMed Holzheimer RG. First results of Lichtenstein hernia repair with ultrapro-mesh as cost saving procedure-quality control combined with a modified quality of life questionnaire (SF- 36) in a series of ambulatory operated patients. Eur J Med Res. 2004;9:323–7.PubMed
15.
Zurück zum Zitat O’Dwyer PJ, Kingsnorth AN, Molloy RG, Small PK, Lammers B, Horeyseck G. Randomized clinical trial assessing impact of a lightweight or heavyweight mesh on chronic pain after inguinal hernia repair. Br J Surg. 2005;92:166–70.CrossRefPubMed O’Dwyer PJ, Kingsnorth AN, Molloy RG, Small PK, Lammers B, Horeyseck G. Randomized clinical trial assessing impact of a lightweight or heavyweight mesh on chronic pain after inguinal hernia repair. Br J Surg. 2005;92:166–70.CrossRefPubMed
16.
Zurück zum Zitat Radford R. Reprocessing single-use devices: choosing a thirdparty reprocessor for a given single-use device. Infect Control Today. 2001;5:27–9. Radford R. Reprocessing single-use devices: choosing a thirdparty reprocessor for a given single-use device. Infect Control Today. 2001;5:27–9.
17.
Zurück zum Zitat Cingi A, Manukyan MN, Gulluoglu BM, Barlas A, Yegen C, Yalin R, et al. Use of resterilized polypropylene mesh in inguinal hernia repair: a prospective, randomized study. J Am Coll Surg. 2005;201:834–40.CrossRefPubMed Cingi A, Manukyan MN, Gulluoglu BM, Barlas A, Yegen C, Yalin R, et al. Use of resterilized polypropylene mesh in inguinal hernia repair: a prospective, randomized study. J Am Coll Surg. 2005;201:834–40.CrossRefPubMed
18.
Zurück zum Zitat Broll R, Bethge T, Windhövel U, Schwandner O, Markert U, Bruch HP, et al. Influence of resterilized polypropylen meshes on growth of human fibroblasts: an experimental in vitro study. Zentralbl Chir. 2002;127:589–93.CrossRefPubMed Broll R, Bethge T, Windhövel U, Schwandner O, Markert U, Bruch HP, et al. Influence of resterilized polypropylen meshes on growth of human fibroblasts: an experimental in vitro study. Zentralbl Chir. 2002;127:589–93.CrossRefPubMed
19.
Zurück zum Zitat Sucullu I, Akin ML, Yitgin S, Filiz AI, Kurt Y. Resterilized mesh in repair of abdominal wall defects in rats. J Invest Surg. 2008;21:171–6.CrossRefPubMed Sucullu I, Akin ML, Yitgin S, Filiz AI, Kurt Y. Resterilized mesh in repair of abdominal wall defects in rats. J Invest Surg. 2008;21:171–6.CrossRefPubMed
20.
Zurück zum Zitat Boscariol MR, Moreira AJ, Mansano RD, Kikuchi IS, Pinto TJ. Sterilization by pure oxygen plasma and by oxygen-hydrogen peroxide plasma: an efficacy study. Int J Pharm. 2008;353:170–5.PubMed Boscariol MR, Moreira AJ, Mansano RD, Kikuchi IS, Pinto TJ. Sterilization by pure oxygen plasma and by oxygen-hydrogen peroxide plasma: an efficacy study. Int J Pharm. 2008;353:170–5.PubMed
21.
Zurück zum Zitat Bathina MN, Mickelsen S, Brooks C, Jaramillo J, Hepton T, Kusumoto FM. Safety and efficacy of hydrogen peroxide plasma sterilization for repeated use of electrophysiology catheters. J Am Coll Cardiol. 1998;32:1384–8.CrossRefPubMed Bathina MN, Mickelsen S, Brooks C, Jaramillo J, Hepton T, Kusumoto FM. Safety and efficacy of hydrogen peroxide plasma sterilization for repeated use of electrophysiology catheters. J Am Coll Cardiol. 1998;32:1384–8.CrossRefPubMed
22.
Zurück zum Zitat Dora CD, Dimarco DS, Zobitz ME, Elliott DS. Time dependent variations in biomechanical properties of cadaveric fascia, porcine dermis, porcine small intestine submucosa, polypropylene mesh and autologous fascia in the rabbit model: implications for sling surgery. J Urol. 2004;171:1970–3.CrossRefPubMed Dora CD, Dimarco DS, Zobitz ME, Elliott DS. Time dependent variations in biomechanical properties of cadaveric fascia, porcine dermis, porcine small intestine submucosa, polypropylene mesh and autologous fascia in the rabbit model: implications for sling surgery. J Urol. 2004;171:1970–3.CrossRefPubMed
23.
Zurück zum Zitat Pernier C, Grosgogeat B, Ponsonnet L, Benay G, Lissac M. Influence of autoclave sterilization on the surface parameters and mechanical properties of six orthodontic wires. Eur J Orthod. 2005;27:72–81.CrossRefPubMed Pernier C, Grosgogeat B, Ponsonnet L, Benay G, Lissac M. Influence of autoclave sterilization on the surface parameters and mechanical properties of six orthodontic wires. Eur J Orthod. 2005;27:72–81.CrossRefPubMed
24.
Zurück zum Zitat Rutala WA, Weber DJ. New disinfection and sterilization methods. Emerg Infect Dis. 2001;7:348–53.CrossRefPubMed Rutala WA, Weber DJ. New disinfection and sterilization methods. Emerg Infect Dis. 2001;7:348–53.CrossRefPubMed
25.
Zurück zum Zitat Rutala WA, Gergen MF, Weber DJ. Comparative evaluation of the sporicidal activity of new low-temperature sterilization technologies: ethylene oxide, 2 plasma sterilization systems, and liquid peracetic acid. Am J Infect Control. 1998;6:393–8.CrossRef Rutala WA, Gergen MF, Weber DJ. Comparative evaluation of the sporicidal activity of new low-temperature sterilization technologies: ethylene oxide, 2 plasma sterilization systems, and liquid peracetic acid. Am J Infect Control. 1998;6:393–8.CrossRef
26.
Zurück zum Zitat Rutala WA, Weber DJ. Clinical effectiveness of low-temperature sterilization technologies. Infect Control Hosp Epidemiol. 1998;19:798–804.CrossRefPubMed Rutala WA, Weber DJ. Clinical effectiveness of low-temperature sterilization technologies. Infect Control Hosp Epidemiol. 1998;19:798–804.CrossRefPubMed
27.
Zurück zum Zitat Rutala WA, Gergen MF, Weber DJ. Sporicidal activity of a new low-temperature sterilization technology: the Sterrad 50 sterilizer. Infect Control Hosp Epidemiol. 1999;20:514–6.CrossRefPubMed Rutala WA, Gergen MF, Weber DJ. Sporicidal activity of a new low-temperature sterilization technology: the Sterrad 50 sterilizer. Infect Control Hosp Epidemiol. 1999;20:514–6.CrossRefPubMed
Metadaten
Titel
For how many times can a partially absorbable monofilament mesh be sterilized? In vitro experimental study
verfasst von
O. V. Ozkan, MD
A. Aydogan, MD
S. Akkucuk, MD
M. Aydin, MD
E. Nacar
B. Ozer
S. Mistikoglu
I. Yetim, MD
A. Gokce, MD
Publikationsdatum
01.04.2013
Verlag
Springer Vienna
Erschienen in
European Surgery / Ausgabe 2/2013
Print ISSN: 1682-8631
Elektronische ISSN: 1682-4016
DOI
https://doi.org/10.1007/s10353-012-0177-5

Weitere Artikel der Ausgabe 2/2013

European Surgery 2/2013 Zur Ausgabe